Instruction manual
30
mounted without any bending stress to the rail, which will cause 
measuring inaccuracies. 
3.  Check that the bolts extending through the rail will not interfere with 
the operation of the moving head. To provide the best accuracy, 
provide large clearance holes in the mounting surface, and shim all 
gaps precisely. 
Notes about Accuracy: 
ProCaliper has been thoroughly tested for accuracy against national 
standards. This test data has been supplied. However, the following 
situations can affect ProCaliper's overall accuracy. The specified 
accuracy of ± 0.05mm (0.002 inches) per measured foot is achieved at 
the heel (rail side) of the jaws. In some cases, this specified level of 
accuracy may be exceeded at the tip of the jaws. This difference results 
from the large distance between the center of the measuring scale inside 
the caliper rail and the tips of the caliper jaws. Small changes in 
temperature can cause slight differences in a given measurement. 
Because ProCaliper's major components are made from aluminum, 
which has a temperature coefficient of 23ppm/°C, it is strongly suggested 
to use ProCaliper in a temperature stable environment (over the 
2400mm [96 inch] measuring range, this coefficient can alter 
measurements by 0.05mm [0.002 inches] per °C). 
Operation: 
1.  Insert the object to be measured between the caliper jaws. While 
holding the object against the measurement face of the fixed jaw, 
slide the right jaw to the other end of the object until it stops. 
2.  Press in and rotate the fine adjustment thumb-wheel until the moving 
jaw is tight against the object. If the position of the moving jaw 
needs to be held in place, use the lock knob on the top of the moving 
head. 
3.  If not using the ProCaliper with SPC data collection devices, press 
the black button on the display to HOLD the displayed value. All of 
the display's operations will be stopped until the black button is 
pressed again. 
4.  Read the measurement data from the ProCaliper's display. The 
displayed data can now  be sent to an SPC data collector/processor 
(if attached) by pressing the black button. 