User`s manual
ProScale M150, M250 with LCD Digital Displays  Page 23 of 48 
SECTION 3    CALIBRATION/MAINTENANCE 
Calibration 
The accuracy of the ProScale system is dependent on the manufacturing of 
the Scale. Since the  Scale is passive, there are  no calibration adjustments 
available or necessary.  
The Displays  have a built-in “scaling factor” that allows correction for slight 
linear inaccuracies that may be observed once ProScale has been installed 
in your application. 
Refer to Section 4: PROGRAMMING (Programming Parameter Pr7) and 
Section 5: ABBE ERROR. 
Once installed, if the direction of movement (+ and  -) on the Display is 
opposite the desired magnitude or direction, the Display programming should 
be changed.  
See Section 4: PROGRAMMING (Programming Parameter Pr0). 
To set a ‘zero’ point or to set in a value representing the current location of 
the ReadHead or measuring point, use the +, 0 and - keys on the Display. 
(See +, 0, and – Keys on page 32). Also see example below.  
The following is an example for calibrating ProScale on a table saw fence. 
Other installations follow the same general procedure. 
1.  Check  to  be sure installation  of  all  parts  is  complete,  all  fasteners are 
secure, and the ReadHead is plugged into the Display. 
2.  Cut a part using the normal fence operation. 
3.  Measure the dimension of the part with the most precise measuring tool 
available (e.g. digital calipers). 
4.  Press  the  zero  key  on  the  ProScale  Display  then  press  and  hold  the 
PLUS key to scroll  until  the  measurement  you  just  made  is displayed. 
(The longer the PLUS key is held down, the faster the Display will scroll.) 
5.  When  the  proper  reading is  reached,  lock the  Display  if  desired.   This 
prevents accidental re-zeroing of the Display.  
For additional information see Section 4: LOCK MODE 