User`s manual
Accurate Technology ProScale  Page 24 of 48 
Compensation Calibration 
This example uses 5 inch measurement standards and an Accurate 
Technology ProTable as the measuring system under calibration. Other 
applications and different length measurement standards would be similar.  
To program the compensation table into the ProScale Display, (firmware 
version must end in C), complete the following steps: 
1.  Be sure that Pr 27 (Accuracy Compensation Enable) is set to 1 and 
that Pr 28 (Compensation Multiple) has been configured to the 
proper length of measurement standards that will be used during 
calibration. See Section 4 PROGRAMMING for instructions on how to 
enter programming mode. 
2.  Turn the Display off. Now press and hold down the zero (0) key and 
the On/Off key for approximately 14 to18 seconds. After that time, 
the display will turn on and complete an LCD segment test. You can 
release the On/Off and zero buttons at this time. 
3.  After the LCD test is complete, the display will show the firmware 
version for about 1 second.  This will be followed by the displaying of 
the current position alternating with the text “no Co”. This indicates 
that the display is attempting to show a position OUTSIDE the 
compensated range.  This is occurring because there is NO 
compensation table data currently programmed into the display.  This 
alternating display will continue throughout the calibration process. 
4.  Place the ProTable moving jaw against the fixed jaw.  Press the zero 
(0) key to zero the current position. 
5.  Press the F1 key. This enters the first compensation point into the 
display.  The display will momentarily show “Co 0” indicating that 
the entry was accepted. 
6.  Open the jaws and place the first measurement standard (5”) against 
the back fence, if installed. Close the moving jaw until the face of the 
jaw gently touches the measurement standard.  Press the F1 key. 
The display will show “Co 1” indicating the first measurement 
standard has been digitized. 
7.  Open the jaws and insert another measurement standard (5”). Close 
the moving jaw until the face of the jaw gently touches the 
measurement standard.  Press F1. The display will show “Co 2” 
briefly.  Continue this procedure with one or more standards placed 
end to end every 5 inches until the maximum measuring range has 
been achieved.  After the last standard distance has been digitized, 
press the F2 key. The display will momentarily show “CoE” 
indicating that the calibration process has ended. 