User`s manual
ProScale M150, M250 with LCD Digital Displays  Page 37 of 48 
Display Functions 
Lock Mode 
The user can “lock-out” the position offset adjustment functions (+, 0, - keys) 
to prevent accidental changes of the current displayed position.  To activate 
the lock mode, press and hold the ON/OFF key and then momentarily press 
the MODE key.  The word LOCK on the LCD Display will turn on or off with 
each lock/unlock operation.  When the LOCK symbol is displayed, the +, 0, 
and - keys will be inactive.  On Displays with an auxiliary keypad: ABS and 
INC modes have independent lock operations.  
(Programming Parameter Pr3) 
Segment Offset Adjustment 
For Scales that are longer than 430mm (17 inches), multiple Scale pattern 
segments are installed end-to-end on the aluminum extrusion. This provides 
a  quasi-absolute  measurement  capability  in  which  the  ReadHead  can 
calculate its position on any individual segment but cannot determine which 
particular segment it is on.  To solve this problem, the Display tracks which 
segment the ReadHead is on by detecting the “joint” between one segment 
and adjacent segments. In certain situations, the crossing from one segment 
to another may not be  detected  by the Display.  This may occur if the 
ReadHead is disconnected from the Display (power source) and then moved 
along the Scale to another segment.  It may also occur if the ReadHead is 
moved too  quickly between  segments.  If the  segment  tracking count  is 
incorrect because of one of the above situations, the user can re-adjust the 
Display to correct the error.  This adjustment is referred to as the SEGMENT 
OFFSET  ADJUSTMENT.    To  add  one  segment  value  (430.08mm)  to  the 
displayed value, press and hold the MODE key and then momentarily press 
the  +  key.    The displayed  position  will increase  by  430.08mm  (16.933 
inches).    To  subtract  one  segment,  430.08mm,  from  the  displayed  value, 
press and hold the MODE key and then momentarily press the - key.   
(Programming Parameter Pr1) 
Offset Addition 
Offset  addition  allows  the  user  to  preset  up  to  3  different  values  that  are 
added to the Display reading when selected.  This function allows the user to 
quickly  switch  from  one reference point  to  another  such  as  in  the  case  of 
inside and outside measurements on the ProPanel measuring system. 
To  enable offset  addition,  programming  parameter  Pr10  must  be  set  to  1.  
The Display will flash 1, 2, 3 or 4 in the upper left corner of the LCD. Offset 1 
is the ABS position with no offset added.  Offset 2 is the ABS position with 
parameter  Pr11  added  to  it.    Offsets  3  and  4  have  similar  functions  with 
parameters Pr12 and Pr13 added to the ABS position respectively. 
To move from Offset 1 to 2, momentarily press the F1 key.  Each depression 
of  the F1 key advances  to  the  next  offset.    After  offset  4, the  Display  will 
move back to offset 1. 
(Programming Parameters 9, 10, 11, 12 and 13) 