User`s manual
Accurate Technology ProScale  Page 38 of 48 
Limit Mode 
The Display will show either “LL“  for Lower Limit or ”UL” for Upper Limit if a 
pre-programmed  upper  or  lower  reading  is  encountered.  Upper  and Lower 
Limits  are  set  with  Programming  Parameters  Pr16  and  Pr17  but  are  only 
active if Pr14 is set to 2. The readout toggles for 2 seconds between current 
position display and "LL" or "UL". This continues as long as limit is exceeded. 
(Programming Parameters 14, 16, 17) 
Caution: Limit monitoring is always active, even in programming mode. 
Scaling 
All  General  Purpose  Displays  have  the  ability  to  “scale”  the  actual 
measurement. This function is useful when the actual measurement must be 
multiplied  or  divided  before  being  displayed. Care  should  be  taken  when 
using this function since  invoking it will cause the  unit to display a reading 
different than the actual measured value. 
(Programming Parameter Pr7) 
Compensation Mode 
When making measurements within the compensated range, the display will 
automatically apply the linear interpolation compensation and display the 
corrected position on the LCD. See Section 3: COMPENSATION CALIBRATION. 
When the ProScale is operated outside the compensated range the LCD will 
alternate between the uncompensated position measurement display and 
“no Co” message. This indicates the ProScale is not within compensation 
range. 
(Programming Parameters Pr27, 28) 
Backlighting (24VDC Displays only) 
LCD backlighting can be configured to always remain OFF, to always remain 
ON or to come on when any key is pressed and then remain on for 3, 7 or 15 
seconds.  
(Programming Parameter 22) 
Monitor Signal Output (24VDC Displays only) 
A  connector  provides  a  solid-state  output  (.1A  /  30  VDC)  that  can be 
configured to  activate on  either a position  drift (MONitor) condition or a 
measurement limit (LIMIT) error.  The output can be configured for normally 
open (N/O, not conducting to ground) or normally closed (N/C, conducting to 
ground).  Although the Display LCD flashes  when  either a position  drift 
(MONitor) condition or a measurement limit (LIMIT) error is encountered, the 
output  signal only  changes  state  once  for  out  of limit  condition  and  then 
toggles back when an in limit condition is restored. Pr14 should be set to 1. 
Caution: These functions are active during Programming. If parameters are 
changed  during  Programming  the  output  signal  could  become  active 
immediately! 
(Programming Parameters 14 and 15) 