User`s manual
Accurate Technology ProScale  Page 40 of 48 
Jumpers 
Surface Mount Displays 
Although the ProScale Display  uses a keyboard-programming  method  to 
enable and  configure features  in the  unit, several selection  jumpers  are 
located on the circuit board for additional system configuration. 
User configurable jumpers consist of three pins and a ‘shorting block’. 
The center  of  these  pins  is  ‘Common’.  One  end  pin  is  labeled ‘A’  and the 
other end pin is labeled ‘B’. 
JP1    FOR FACTORY USE ONLY 
JP2    Absolute/Incremental ReadHead Selection 
The General Purpose (Battery operated) Displays support both Incremental and 
Absolute style measuring systems (See Section 1: Terminology). To configure the 
Display for use with absolute ReadHeads (default), install the shorting jumper in 
position A. For incremental ReadHeads, install in position B. For Lithium Battery use 
install on position A. 
JP3    Programming Enable/Disable 
Front panel entry to the programming mode of the Display can be enabled or 
disabled. To enable programming (default), install the shorting jumper in position A. 
To disable programming, install in position B. When programming mode is disabled, 
the user cannot access the programming functions via the Mode + 0 keys as 
described in the Section 4: PROGRAMMING. This provides the user with a method of 
configuring the Display with specific parameters and prevents unauthorized 
configuration changes. 
JP4    Display Power 
This jumper configures the Display to operate on either Battery or 24VDC. 
This jumper will be set at the factory based on the type of Display you have ordered. 
24VDC Displays are not compatible with INCremental systems. 
JP5    Battery Selection 
1.5v Alkaline Battery: jumper must be placed across B pins 
3.6V Lithium Battery: jumpers must be placed across A-B and B-C (see also JP2) 
JP1
JP2
JP3
JP4
JP5
Surface 
Mount 
Display 
Circuit 
Board  