User's Manual

Quality Control of the System
16
Using Glucose Control Solution (Sold Separately)
The purpose of ACEZIN GLUGold Control Solution is to perform a quality
control test to verify whether your meter and test strips are working
together properly.
There are two levels of Glucose Control Solutions for different acceptable
ranges: High Level and Medium Level. We recommend testing with both
levels of Glucose Control Solution.
How does it work?
A known amount of glucose in the control solution reacts with AMG011
Blood Glucose Test Strip.
Who or when to perform a control solution test?
Handling your meter at the first time.
After opening a new vial of test strips.
Strong impact on the meter.
When your symptoms do not agree with the way you feel, or when the result
is higher or lower than expectation.
Whenever users wish to check whether Blood Glucose Test Strip and Blood
Glucose Meter are working properly.
1
Insert test strip into test strip slot.
When symbol flashes, press S button to activate
Quality Control Mode.
With symbol on, your meter is ready to perform
quality control test.
Note
For accurate results on both blood glucose measurements and quality
control tests, use ACEZIN GLUGold Control Solution only.
Results of quality control tests will not be stored in the memory.