Quality Control of the System
17
3
Shake the vial of Glucose Control Solution well
before use. Squeeze the vial but discard the first
droplet.
Squeeze the second droplet to your clean fingertip or on
a clean object such as a transparency.
Move the collection zone of test strip close to Glucose
Control Solution droplet. Collection zone automatically
drink the droplet in. When full, your meter beeps and
counts down for 5 seconds.
4
When results match with acceptable range
printed on
the Test Strip vial, your System is working properly.
NOTE
To avoid contamination,
do not apply the control
solution to the test strip
directly from the bottle.
Note
Check the Test Strip vial for acceptable ranges for High
and Medium Levels. Refer to Glucose Control Solution
instruction before use.
Understanding Your Test Results
When results match with acceptable range printed on the Test Strip vial,
your System is working properly, and your test results are reliable.
If your results do not match with the acceptable range, please check the
following:
Are the control solutions or test strips within the expiration date?
Has the control solution been contaminated?
Did you follow the test procedures correctly?
Please repeat the test. If results still do not match, contact your local Ace
Medical office or the authorized distributor.