2.4. Lighting protection and Earthing
To protect people and the equipment from lighting conditions, it is mandatory to observe the
following:
Always install the equipment in a place protected by a lighting rod.
Always ground the equipment using the dedicated M4 tapped hole.
Secure the M4 screw with 1.2 Nm maximum torque.
Ground the equipment with an AWG8 or 10 mm
2
cable.
Ground the equipment at the shortest distance.
Plan periodic reviews of the grounding connection and correct immediately any
damages.
In case of doubt, refer to qualified people before performing any installation or work on
the equipment.
Use Panduit P8-8RN-T non isolated ring tongue or equivalent.
The operator need to use a Torx screw driver and the crimping tool for the ring tongue (these
tools are not supplied with the equipement).
Figure 1: ATD-300/ATD-301 grounding thread
2.5. Operating Environment