User's Manual
Don't use this instrument, and its accessory in an explosive atmosphere, or near flammable
products, or near corrosive products.
Warning: Take care to use the instrument in accordance with the environmental
constraints
2.6. Installation Caution
The installation must be done only by workers qualified and authorized to work at a high level.
The equipment is fixed in height, it is the responsibility of the contractor to ensure the safety of
the installation. For a safe installation the requirements below must be followed:
● Keep the work area clean and well lit. Cluttered or dark areas invite accidents.
● Ensure the cables connected to the instrument are secure, protected and fixed to avoid
hanging on or falling.
● Before any use, visually inspect the instrument case and its accessories and check there
is no visible damage such as crack, broken part or abnormal deformation.
● Do not install nor uninstall all or part of the equipment in stormy weather.
● Use the adapted Personal Protective Equipment and fall protection to work in high level
● Adapt fixation to the weight of the equipment (including external load: wind, snow...) and
to the material of the wall
● Control the strength of the attachment point
● Use thread lock on fixation to prevent self-loosening
2.7. Electrical Safety
● All external circuits connected to the device should be non-hazardous voltage,
and energy source limited within the meaning of the IEC61010-1 standard.
● The PoE power supply (not provided) must follow the IEEE 802.3af / 802.3at standard.
● Do not abuse the cord. Never use the cord for carrying, pulling or unplugging the
instrument.
● Ensure the PoE power supply is connected to an outlet protected from shocks or
possible damage.
● Keep cord away from heat, oil, sharp edges or moving parts. Damaged or entangled
cords increase the risk of electric shock.
ACOEM | ATD-300/ATD-301 | User Manual FW 1.0.0.4 8