1.14.7 Pyrometer
For the pyrometer please read the safety instructions delivered with the instrument (printed and on the CDROM)
From any screen, read the temperature with the built-in pyrometer
Notes:
from the Collect module, measurement list screen: if in Measurement setting, if “Pyrometer” = “Internal”,
temperature measurement of the route with input type = DC will be done with the built-in pyrometer
The pyrometer measures the average temperature in a circle. The diameter of the circle depends on the
distance to the target. The diameter is about 4 cm at a distance of 50 cm.
Warning: due to the distance between the pyrometer cell and the laser source, the circle is decentred by 3
cm.
1.14.8 Stroboscope
For the stroboscope please see the safety instructions delivered with the instrument (printed and on the CDROM)
From any screen, read the rotation speed with the stroboscope
From the Collect module, measurement list screen: check or adjust rotation speed of the machine.
From the Collect module, spectrum display screen: check or read rotation speed of the machine.
Fine tune. A continuous press speeds up the modification
Fast tune. A continuous press speeds up the modification
Divide or Multiply the value by 2.
Set the flash duration: default value is 5 degrees. The longer the duration, the brighter the flash, but
the fuzzier the target on the rotor. Limits range from 0.5 to 15 degrees.
This function is only accessible in Collect module.
For machine with variable rotation speed: rotation speed measurement of the machine is updated by the
stroboscope value.
For machine with fixed rotation speed: rotation speed setting of the machine is updated by the stroboscope
value. In this case, when downloading the route, the new rotation speed is used to update the NEST
initial setting
Notes:
It is recommended to start with a frequency higher than the rotation and decrease gradually until stopping
the rotor marker.
Default unit is set in "Spectrum display" parameters (see § 2.5)
Once the marker is stopped, to be sure not to be on a sub-multiple of the rotation, use , the marker
should appear 2 times. Use to return to the initial frequency.
1.14.9 Screenshot
From anywhere, you can save a screen copy. Images are stored in folder “Screenshots”. Connect your PC to
FALCON to copy them (see § 1.12).