User's Manual

31
Brand of ACOEM
2.11 DATA MANAGEMENT
Used to delete all data of a particular module (Collect or Balancing) or clear all the data from the instrument.
Reset collect: delete all data of Collect module
Reset balancing: delete all data of Balancing module
Full reset: delete all data of the instrument
Reset setup: return to instrument initial configuration
Data are definitively deleted.
Export log: create an event log file in the ‘Export’ folder. This file can be used by Acoem support for
troubleshooting.
2.12 UPDATE FIRMWARE
It is first recommended to make a backup of the instrument memory.
For more details, see § 5.4
Put update firmware (.czip file) on a USB memory key.
Notes:
o The czip file must be at the root of the USB memory key
o It must have only one update file.
o USB memory stick format must be FAT32, NTFS format is not accepted. It is always possible to
reformat it to FAT32
Connect the system to its power supply.
Plug the USB memory key in FALCON connector G.
Shortcuts > Setting > Update firmware
Click on “Read USB memory”.
Once the new firmware is detected click on “Update firmware”.
Wait until the system restart.
2.13 AUTO TEST
Run tests on the main components of the instrument. This operation takes about 3 mn. To check the WLS sensor,
it is first necessary to connect it to the instrument (see § 1.11.2).