User's Manual

60
Brand of ACOEM
3.7.6 How to import templates in FALCON
New machines or new measurements can be added in the OFF_ROUTE route using point or measurement
template from OFF_ROUTE route or any other route stored in the instrument.
A template route with usual measurement examples is supplied on the CDROM in “OFF_ROUTE TEMPLATES”
folder. It is available in 2 formats:
Templates_EN.zld: this route can be imported directly in the instrument:
o Copy the file:
On a folder “Import” of a USB memory stick
or
Directly in the folder “Import” of the instrument memory (see § 6.1)
o Go to “Collect” module
o If you are using a USB memory stick, insert it in the port G behind the trapdoor
o Click on Import
o Select the route to import.
Templates_EN.zip: this file can be imported in XPR (see XPR user manual § 11.4.2). In XPR each
measurement properties can be adjusted and then loaded to the instrument as usual machines.
The document “Templates-EN.pdf” gives a description of the route content.
3.7.7 Download OFF_ROUTE measurements to the PC
Proceed as for a standard route: see § Erreur ! Source du renvoi introuvable. When the route is downloaded,
machines are created in a specific location named “OffRoute of the active database. They can be used as new
machines or the measurements can be merged with existing machines. For more details, see NEST documentation.
Note: limitations for measurement properties downloaded to the PC
Labels of “Other parameter” and “Other unit” are not managed in Falcon and are replaced by-“ character.
“Velocimeter IEPE” input type is not managed on PC database and it is replaced by “AC-V” input type.