User's Manual

75
Brand of ACOEM
MAINTENANCE CHAPTER 5.
5.1 RESET
Open the connector trapdoor. The reset button is in front of the small hole. Press it with a thin tip (a paperclip for
example) during 5 or 6 seconds. The instrument restarts on the home screen. No data are deleted with this
operation.
5.2 CLEANING
To clean up the instrument, use only soft products (lightly soaped water). For delicate surfaces (screen, pyrometer,
tachometer and laser lenses), use only clean compressed air to remove particles. If needed, gently wipe the
surface using a soft, damp cloth.
5.3 CALIBRATION
See § 2.14.
5.4 BACKUP OF INSTRUMENT MEMORY
Connect first the instrument to the PC (see § 1.12). You can now copy the data from the instrument to your PC for
backup purpose. The detail of the organization is explained in Appendix 1 (see § 6.1).
5.5 INSTRUMENT FIRMWARE UPDATE
See § 2.12.
5.6 WLS SENSOR FIRMWARE UPDATE
Switch on the WLS sensor: press on/off for 2s
Connect it to a PC using the USB cable
Wait until its drive is visible from the PC
Copy the new firmware file (.czip file) in the ‘Update’ folder
Don’t disconnect the USB cable until the end of the update
Switch off the sensor: press on/off for 6s
Switch on the sensor, the red LED is flashing during the update (about 30 s).
When the update is completed, the blue LED turns on. The .czip file is deleted. A report file
update_report_x.xx.txt, where x.xx is the version number, is located in the root. The content of the file is
‘UpdateDone: x’ with x=0 in case of an error.