6.2.2 Long time-wave measurement (option)
The following tables indicate the maximum number of samples and time wave duration according to:
Number of simultaneous channels
Analysis frequency
There are two cases to consider:
Each point has only 1 time wave in the measurement list (1)
There is also an overall level measurement in the list (2)
Depending of the configuration, if the time wave is displayed during the acquisition, the limit can be
extended to the one of the previous case.
(1) Only one time wave measurement per channel
Analysis/sampling
frequency
Max number of samples / acquisition time (s)
1 channel
2 channels
3 channels
4 channels
80 / 204.8 kHz
4M / 20 s
2M / 10 s
40 /102.4 kHz
4M / 40 s
2M / 20 s
1M / 10 s
1M / 10 s
20 / 51.2 kHz
4M / 80 s
2M / 40 s
1M / 20 s
1M / 20 s
10 / 25.6 kHz
4M / 160 s
2M / 80 s
1M / 40 s
1M / 40 s
5 / 12.8 kHz
4M / 320 s
2M / 160 s
1M / 80 s
1M / 80 s
2 / 5.12 kHz
4M / 800 s
2M / 400 s
1M / 200 s
1M / 200 s
(2) Time wave measurement with an overall level measurement in the list
Analysis/sampling
frequency
Max number of samples / acquisition time (s)
1 channel
2 channels
3 channels
4 channels
80 / 204.8 kHz
2M / 10 s
2M / 10 s
40 /102.4 kHz
2M / 20 s
2M / 20 s
1M / 10 s
1M / 10 s
20 / 51.2 kHz
2M / 40 s
2M / 40 s
1M / 20 s
1M / 20 s
10 / 25.6 kHz
1M / 40 s
1M / 40 s
1M / 40 s
1M / 40 s
5 / 12.8 kHz
512K / 40 s
512K / 40 s
512K / 40 s
512K / 40 s
2 / 5.12 kHz
128K / 25 s
128K / 25 s
128K / 25 s
128K / 25 s