2. SENSORS AND EXPANDERS
2.1. IDENTIFICATION
Table 2: sensors and expander part numbers
Model
Device
EGL1102000
Uni-axial sensor
EGL1103000
Tri-axial sensor
EGL1104000
Expander
2.2. INSTALLATION / DISASSEMBLY
During installation or disassembly of Eagle devices, the operator must wear the prescribed
PPE (Personal Protective Equipment), respect the regulations in force and have the
necessary clearances.
WARNING: In hazardous environments, the operator has to arrange organizational safety
measures which reliably prevent the occurrence of an ignitable atmosphere, by default
reduce the probability that a flammable atmosphere can occur at all (employing the use of
suitable gas detection systems).
A damp cloth should be used when installing or taking off the sensors and expanders in all
hazardous areas to eliminate static electricity resulting from this operation.
Only use the dedicated tube wrench to fasten or unfasten sensor/expander.
Only use your hands or a rubber strap wrench to remove the protection tube from the
sensor/expander. The use of tongue-and-groove pliers (also known as water pump pliers,
adjustable pliers, groove-joint pliers, arc-joint pliers, Multi-Grips, and Channellocks) are
forbidden and will cause damages and lose of sealing IP67.
CAUTION:
Do not grasp the sensor by the plastic cap.
Do not hit the plastic cap.
Do not tighten the sensor by the plastic cap
Figure 1 : sensor installation tools