Owner's manual

CALMASTER
Operating Instructions
www.actmeters.com
6
www.actmeters.com
3
CALMASTER
Operating Instructions
Contents
First Steps ..................................................................................................... 4
Testing AC Voltage ....................................................................................... 4
Testing Resistance ....................................................................................... 5
Testing DC Voltage ..................................................................................... 5
Testing DC Current ...................................................................................... 6
Specications ............................................................................................... 6
Testing DC Current
1. Connect ‘Calibrated Meter’ and ‘Meter Under Test’ to the CALMASTER
right sockets
2. Switch both meter to ‘DC mA’ range
3. Conrm that the ‘Meter Under Test’ mA reading is within ±5% of the
reading displayed on the ‘Calibrated Meter’
4. Log readings obtained for calibration records
5. Repeat above on all DC current ranges on the meters
WARNING: ON 10AMP RANGE MAX TEST TIME 30 SECONDS
Fuse Protection (AC Input) 250VAC, 1.6Amp max
(AC Output) 250VAC, 50mA max
Weight 2.2KG
Dimensions 216(L) x 132(W) x 102(D)mm
Resistance AC Volts DC Volts DC Current
00.0Ω up to 100K 0.0 to 230VAC 0.0 to 36VDC 1mA
10mA
100mA
1Amp
Specications