Technical information

Technical Information 2
ECD Management in EMEA and IDO Countries 17
4 Cap the inlet and outlet fittings when the ECD is not in use.
5 Connect the ECD exhaust vent to a fume hood or vent to the outside in
some other acceptable manner whenever the ECD is in use.
ECD warnings
Although beta particles at this energy level have little penetrating power (the
surface layer of the skin or a few sheets of paper will stop most of them), they
may be hazardous if the isotope is ingested or inhaled. For this reason, the cell
must be handled with care, radioactive leak tests must be performed at
intervals required by your local Nuclear Agency, the inlet and outlet fittings
must be capped when the ECD is not in use, corrosive chemicals must not be
introduced into the detector, and the detector must be properly vented outside
the laboratory environment.
WARNING
Materials that may react with the
63
Ni source, either to form volatile products or to
cause physical degradation of the plated film, must be avoided. These materials
include oxidizing compounds, acids, wet halogens, wet nitric acid, ammonium
hydroxide, hydrogen sulfide, PCPs and carbon monoxide. This list is not exhaustive
but indicates the kinds of compounds that may cause damage to
63
Ni detectors.
WARNING
In the extremely unlikely event that both the oven and the ECD heated zone should go
into thermal runaway (maximum, uncontrolled heating in excess of 400 °C) at the
same time and that the ECD remains exposed to this condition for more than
12 hours, the following must be done:·
After turning off the main power and allowing the instrument to cool, cap the ECD
inlet and exhaust vent openings. Wear disposable plastic gloves and observe
normal safety precautions.
Return the cell to Agilent Technologies following instructions included in this
booklet (see Chapter3).
It is unlikely, even in this very unusual situation, that radioactive material will
escape from the cell. However, permanent damage to the
63
Ni plating within the cell
is possible and, therefore, the cell must be returned for exchange.