Instruction Manual

2-Pole Interferential (Premodulated)
2-Pole Interferential Waveform
Waveform Type: Amplitude modulated sine wave
Polarity: None
Maximum Voltage: 50 V (Peak value, 500Ω load)
Current: 0–100 mA peak, 500Ω load
Carrier Frequency: 2 kHz, 4 kHz, 5 kHz
Interference frequency 1~250 beats
Frequency Modulation: 1 to 10 in 1-beat steps, and
10 to 250 in 10-beat steps.
Available Channels: All
EMS Mode
EMS Waveform
Waveform Type: Amplitude modulated sine wave
Polarity: None
Maximum Voltage: 50 V (Peak value, 500Ω load)
Current: 0–100 mA peak, 500Ω load
Carrier Frequency: 2 kHz, 4 kHz, 5 kHz
Interference frequency 20 to 250 beats in 10 beat steps
Amplitude Modulation
Options: Independent, Simultaneous, Alternate
On-time 1 to 30 sec (In 1-sec steps)
Off-time 1 to 99 sec (In 1-sec steps)
Ramp-up Time 0 to 3 sec (In 0.5-sec steps)
Hold Time 0 to 30 sec (In 0.5-sec steps)
Ramp-down time 0 to 3 sec (in 0.5-sec steps)
Contraction 1 to 999 times
Available Channels: All
Russian Mode
Russian Waveform
Waveform Type: Burst modulated sine wave
Polarity: None
Maximum Voltage: 50 V (Peak value, 500Ω load)
Current: 0–100 mA peak, 500Ω load
Frequency: 2500 Hz, Burst at 10 ms on and 10 ms off
Available Amplitude
Modulation Options: Independent, Simultaneous, alternate
On-time 1 to 30 sec (In 1-sec steps)
Off-time 1 to 99 sec (In 1-sec steps)
Ramp-up Time 0 to 3 sec (In 0.5-sec steps)
Hold Time 0 to 30 sec (In 0.5-sec steps)
Ramp-down time 0 to 3 sec (in 0.5-sec steps)
Contraction 1 to 999 times
Available Channels: All
Hi Volt Mode
Hi Volt Waveform
Waveform Type: Monophasic twin peak
Polarity: Positive, negative or both
Voltage: 0 to 150 V (Peak value, 500Ω load)
Maximum Current: 300 mA peak, (Peak value, 500Ω load)
Phase Duration: 10 to 80 μs (In 10-μs steps)
Frequency: Constant0.5 to 200 Hz
Burst: 60 Hz (Fixed)
Sweep1 to 200 Hz
Independent, Simultaneous,
Alternate: 20 to 200 Hz
Frequency Modulation: Constant, Burst, Sweep
Burst frequency *2 0.5 to 7 Hz (0.5, 0.7 Hz,
1 to 7 Hz (In 1-Hz steps)).
Available Amplitude
Modulation Options: Independent, Simultaneous, Alternate
On-time *1 1 to 30 sec (In 1-sec steps)
Off-time *1 1 to 99 sec (In 1-sec steps)
Ramp-up time *1 0 to 3 sec (In 0.5-sec steps)