Owner's manual
3
Time
Output
1, 3.2 or 3.3 MHz
Ultrasound
Pulse Width
2 ms
Pulse Space
10 ms
Figure 3.1—Pulse Waveform—20% Duty Cycle
Time
Output
1, 3.2 or 3.3 MHz Ultrasound
Pulse Width
5 ms
10 ms
Pulse Space
Figure 3.2—Pulse Waveform—50% Duty Cycle
In the continuous mode, the power is on at least 95% of the
time the timer is running. The continuous mode waveform is
shown below:
Time
Output
1, 3.2 or 3.3 MHz Ultrasound
Figure 3.3—Continuous Waveform
Ultrasonic Applicator Specifications:
Piezoelectric discs: The output transducer utilizes a barium titanate disc with a
specially coated face.
Individual Applicator Specifications:
Applicator Part Number Frequency Effective Radiating Area
ME 7310 1 MHz ±5% 10 cm²±10%
ME 7331 3.3 MHz ±5% 1 cm² ±10%
ME 7513 1 or 3.2 MHz ±5% 5 cm² ±10%
Maximum Beam
Non–Uniformity Ratio:
6:1
Maximum Effective
Intensity Ratio:
2:1
Spatial Pattern: The applicator produces a collimated (cylindrical) beam with
an area of 1, 5 or 10 cm
²
, measured 5 mm from the ceramic
disc surface when the radiation is emitted into the equivalent
of an infinite medium of distilled water at 30° C.
The beam of the applicator is circular in all planes parallel to