Datasheet

Gravity as a Calibration Reference
Acceleration can be measured in units of gravitational force or "G", where 1G represents the
gravitational pull at the surface of the earth. Gravity is a relatively stable force and makes a
convenient and reliable calibration reference for surface-dwelling earthlings.
Calibration Method:
To calibrate the sensor to the gravitational reference, you need to determine the sensor
output for each axis when it is precisely aligned with the axis of gravitational pull. Laboratory
quality calibration uses precision positioning jigs. The method described here is simple and
gives surprisingly good results with just a block of wood.
Mount the Sensor:
FIrst mount the sensor securely to a block or a box. The size is not important, as long as all
© Adafruit Industries https://learn.adafruit.com/adxl345-digital-accelerometer Page 11 of 17