Specifications

37
| P a g e © Adam Equipment Company 2011
The parameters available are:
F4 Int Initial Zero Range
F5 rEZ Re-Zero range
F6 SCS Successive Tare Enable
F7 Cnt Display the A/D counts
F8 Zem Zero Mode
F9 Lvd Low voltage detection
FA AZn Auto Zero Range
Fb FPS User Function Password
The description of the parameters is shown in section 12.3
12.2 USING
2006
TO
ENTER
THE
SECURE
PARAMETERS
Press the [Z/T] key during the display counting when turned on,
When Pn is displayed, enter the number 2006 using the [Unit/] and
[Pcs/] keys and then press [Z/T]. The displays will show the first
parameters, called F1 CAL”.
To select another parameter press the [Pcs/] key to advance through the
parameters.
Press the [Z/T] key to enter a parameter.
To exit a parameter, press the [Print/M+/Esc] key.
The display will show the parameter number and a name.
When a parameter is entered by pressing the [Z/T] key, the displays will
guide you through the parameter selected and the options available.
The parameters available are:
F1 CAL To enter the Calibration
F2 dEC Decimal Point Position
F3 CAP Default Weighing Unit and capacity
F4 Int Initial Zero Range
F5 rEZ Re-Zero range
F6 SCS Successive Tare Enable
F7 Cnt Display the A/D counts
F8 Zem Zero Mode
F9 Lvd Low voltage detection
FA AZn Auto Zero Range
Fb FPS User Function Password