CODE
Input new password. If user does not want to modify
password, press [PRINT] to enter the next parameter.
P1 FIL
{ Filter }
F1 FIL
Filtering. Pressing the M+ key increases the intensity
of the filter, the higher the filter value the more stable
the weight reading is.
F2 ZEO
Zero track parameter. If the zero point often moves,
user can set this parameter. Pressing the M+ key
increases the zero tracking filtering, however
increasing this value may make it difficult to weigh
small items.
F3 ZTR
Creep tracking parameter. If the connected load cell
suffers from creep, user can set the parameter.
Pressing the M+ key increases the creep tracking
filtering, however increasing this value may make it
difficult to weigh small items.
P2 C-D
{Capacity and
division}
The unit weight window :
CXXXX
XXXX is the capacity. If weigh is more than
capacity+9 d, display will shows over load
The total quantity
window : DXXXX
Division value.
dp xxx Open or close the dual division.
CU xx The unit of capacity and calibration.
P3 LIN X
{ Multi-range
calibration }
User can set multi-range calibration on/off here. NoteIf user does not want to
use the multi-range calibration, please set X off. If choose on, please remove all
the weights from the pan before doing multi-range calibration.
LINEX
The state of multi-range calibration, X can be 1, 2, 3,
it means setting step. Input the weights value, the put
the right weights to calibrate.
Remark: In the second step, value of weights must
bigger than the first step, and in the third step, value
of weights must bigger than the second step.
P4 CAL
The total quantity
window : WEIXXXX
Use number key to input the weight value, XXXX is
the value of the weights for calibration. Press [PLU] to
enter the calibration. The display will show the A/D
value, with no weight on the top pan press the PLU
key when the stable indicator is showing
The unit weight window :
LOAD
The total quantity
window : XXXXX
In this step, put on the correct weight, after stable
indicator appear, press [PLU] can complete
calibration.