8.0 CALIBRATION
8.1 SETTING CAPACITY AND DIVISION
To set up the capacity and division follow the procedure below:
With the indicator turned off, press and hold down the [PLU] and [CE/Backlight]
keys and then turn the scale on using the rocker switch at the right back hand side
of the indicator. The display will show:
Enter 123456 which is the default passcode and you should see the following
screen:
Press the [Print/Save] button, this will take you into the system parameters (P1 FIL
is to do with filtering) , press the [Print/Save] button again, and you should see the
system parameter for settings capacity & divisions (P2 C-d), screen should look
similar to as follows: