In the above example you can see the capacity is set for 20kg x 1g.
To increase the capacity by 10 press the [M+] key; the capacity reaches a
maximum by continuing pressing the [M+] key and then returns to the lowest point.
Pressing the [Zero/0] key increases the flashing capacity value to various pre-
set values, pressing the [Tare] key decreases the flashing capacity value.
Once you are happy with the capacity value being correct, press the PLU key.The
indicator has fixed divisions that can be set, but these are all normal multiples.To
change the flashing division value press the [M+] key, repeated pressing of the M+
key cycles the values round again.
Once you are happy with the division, press the PLU key, you will then see the
following screen:
The flashing dp, refers to dual division; leave this set to off and press the PLU
button to move to setting calibration unit which is shown as CU kg this is the
weighing unit for calibration. To change the weighing unit form kg to lb press the
[M+] key.
To move to the next parameter group setting press [Print/Save] key. The
parameter shopwn on the display is P3 LIN (muli range calibration) leave this set
to OFF; if it is set to ON pressing the [M+] key toggles the setting.
Press the [Print/Save] key to go to calibration -P4 CAL