12.0 ERROR CODES
During the initial power-on testing or during operation, the scale may show an error
message. The meaning of the error messages is described below.
If an error message is shown, repeat the step that caused the message, turning the
balance on, carry out the calibration or other functions. If the error message is still shown
contact your dealer for further support.
ERROR CODE DESCRIPTION Solution
ADO---
Over A/D resolution
range
Remove the weights of the pan or send
back to the agent.
ADL---
Over A/D resolution
ran
g
e
Remove the weights of the pan or send
back to the a
g
ent.
OVR---
Over load (max: capacity
+9e
)
Check the weight of the weights,
cannot exceed Max Cap.+9e
ERR-Z
AD value exceed the pow
er on zero range when po
we
r
on
Power on again after remove the
weights or re-calibrate
Battery symbol
flash
Low power alarm Charge battery
Lo-Bat
Low power alarm Charge battery