3.0 INSTALLATION
3.1 UNPACKING
This scale must be connected to a load cell platform and calibrated as necessary to match
the platform and user requirements.
The user’s application and the technical specifications of the platform or load cell will
determine the necessary configuration.
3.2 LOCATING
The scale should not be placed in a location that will
reduce the accuracy.
Avoid extremes of temperature. Do not place in
direct sunlight or near air conditioning vents.
Avoid unsuitable tables. The table or floor must be
rigid and not vibrate.
Avoid unstable power sources. Do not use near
large users of electricity such as welding equipment
or large motors.
Do not place near vibrating machinery.
Avoid high humidity that might cause condensation.
Avoid direct contact with water. Do not spray or
immerse the indicator in water.
Avoid air movement such as from fans or opening
doors. Do not place near open windows or air-
conditioning vents.
Keep the scale clean. Do not stack material on the
indicator when they are not in use.