Product Manual
© Adam Equipment Company 2019 12
Mass on the platform is above the high limit
Mass is between the limits
Mass is below the low limit
6.5.1 Setting up check weighing parameters
In normal weighing mode, hold down [Mode/Print] to enter the check weighing
function.
The limits are displayed in the weighing unit in use.
The display will show “– H – L -”. The hyphens represent the audible alarm whilst
the “H” represents the high limit and the “L” represents the low limit. The alarm can
also be set to OFF.
When the displayed hyphens flash, this is an indication that the beep is ON.
There are 6 various customisable alarm options:
- H L
The beep is on when the weighing mass is more than the set high limit.
H - L
The beep is on when the weighing mass is between the set high and low limit.
H L -
The beep is on when the weighing mass is below the set low limit.
- H L -
The beep is on when the weighing mass is both more than and below the set
limits.
- H - L
The beep is on when the weighing mass is in between the set limits and more
than the set high limit.
H - L -
The beep is on when the weighing mass is in between the set limits and below the
set low limit.