Product Manual

© Adam Equipment Company 2019 15
To use the Animal Weighing function it is necessary to set the amount of filtering
required for the animals to be weighed. More active animals will require a higher
level of filtering to give a stable result.
The display will show XXd - Hdwhere x is a value from 10 d to 70 d. The higher
the value the greater the amount of filter there will be. To increment the value
shown press the [Units] key, to decrease the value press [Zero]. Press the
[Mode/Print] key to accept it. Press [MR] twice, then press [Zero] to return to
weighing mode.
The indicator is now ready to weigh in Animal weighing.
6.7.1 Animal weighing procedure
With the platform empty the display will show zero weight. Place containers or
blankets onto the platform and press the [Tare] key to zero the display. The
indicator may go into the animal weighing procedure when the items are placed on
the platform but will return to showing zero when the [Tare] key is pressed.
Place the animal to be weighed on the platform.
When a stable reading is found, the display will show this value, and the display will
temporarily lock. A beep and padlock symbol will appear on the display to indicate a
stable reading has been found.
Manual unlocking and Peak hold options are available within the animal weighing
function. Enter the animal weighing function as instructed in the above section.
To select the manual unlocking option within animal weighing, the display will
show XXd - Hd. Use the [Tare] key until the display shows “MNU”. This is
the Manual unlocking option within the animal weighing function. This will
hold the stable value and can be manually unlocked using the zero key.
Press [MR] twice and then press [Zero] to return to normal weighing.