7.0 CALIBRATION
The indicator can be calibrated using a range of different unit weights, depending
on the weighing unit selected at the time of calibration.
Calibration before use is recommended for best accuracy. It is suggested that the
indicator be calibrated when first installed and any time the indicator is moved to a
different environment.
7.1 SIMPLE CALIBRATION
To start the calibration, in normal weighing mode ensure all weights are removed
from the platform. Hold down [Tare] for 3 seconds to enter the calibration function,
the display will show “-CAL-” and then the calibration weight value.
Enter the desired calibration weight value using [Units] to move the flashing digit,
[Tare] and [Zero] to increase or decrease the value, the display will now show the
calibration mass expected.
Place the weights onto the platform, when the stable symbol is displayed press the
[Mode/Print] key to confirm the calibration. The display will then return to normal
weighing and display the calibrated result.
NOTE:
If the calibration result is outside the calibration value selected by more than double or less
than half then the calibration cannot pass.