Product Manual
© Adam Equipment Company 2019 27
9.2 LABEL DOWNLOAD PROCEDURE VIA RS 232 CABLE
Save your final label design in preparation for download via RS 232 to the indicator
by pressing the save icon located in the top left hand corner of the software.
Export the image by selecting the export icon from the barcode icon in the top left-
hand corner
Press the download icon
Select the necessary serial port and baud rate and click “download”.
10.0 PARAMETER SETTINGS
Pressing the [M+] key allows the user to access the parameters for customising the
indicator. The parameters are split into 3 groups-
1. Unit parameters (P1 Unt)
2. Communication parameters (P2 trA)
3. Indicator Functions parameters (P3 FUN)
When [M+] is held down for 3 seconds, the display will first show “P1 Unt” for Unit
parameters.
Press the [MR] key to advance through the groups “P1 Unt”, “P2 trA” , “P3 FUN”.
Press [Mode/Print] to enter the desired group of parameters.
If you press [Zero], the indicator will exit the User Parameter section and return to
normal weighing.