Product Manual
© Adam Equipment Company 2019 37
12.1.3 Multistage calibration (P6 LI)
Parameter Description Option
LINE 1 “LINE 1” will appear briefly on the
display to show that you are entering
the linearity calibration.
XXXXXX Set the 1
st
linearity calibration weight
value; place the weight on the scale,
press [MODE/PRINT].
Use [Units] key to move flashing
digit, [Tare] to increase and
[Zero] to decrease the value.
LINE 2
Briefly
displayed
Set the 2
nd
linearity calibration weight
value; place the weight on the scale,
press [MODE/PRINT].
LINE 3
Briefly
displayed
Set the 3
rd
linearity calibration weight
value, place this weight on the scale,
press [MODE/PRINT]. Display shows
‘good’ if linearity successful
Scale automatically then steps on to next parameters P7 (CAL).
12.1.4 Calibration (P7 CALl)
Parameter Description Option
XXXX WEI Enter the weight of the weight that will
be calibrated.
Use [Units] key to move flashing
digit, [Tare] to increase and
[Zero] to decrease the numeric
values. Press [Mode/print] to
move onto the next step.
XXXXX CAL AD value of zero, the platform must
be empty.
Wait for the stabilization symbol to
appear.
Once stabilized press
[Mode/print] to move onto the
next step.
XXXXX LOD AD value, at this time place the
previously entered weight on the
platform.
Wait for the stabilization symbol to
appear.
Press [Mode/print] to calibrate, if
calibration is successful “PASS”
will be displayed.
At this point the indicator returns
to normal weighing mode