12.1.5 Calibration store and restore setting (P8 EC)
Parameter Description Option
E1 ruL “E1 ruL” will display for 1 second,
XXXX is the calibration value.
For example, if you do not have the
suitable weights by entering the
calibration value this can then be
restored.
Press [Units] to modify the
calibration value and move the
flashing digit.
Press [Tare] to increase and
[Zero] to decrease the numeric
values then press [Mode/Print]
to move onto the next setting.
E2 Ldo “Ldo” will display for 1 second,
XXXX is the value of multistage
calibration/linearity.
It has 4 parameters, if user has
carried out multistage calibration then
the parameters are effective. The
values can be enter which means
multi stage calibration does not need
to be done.
The setting will go through the
various parameters of the
multistage calibration.
Press [Units] to modify the
calibration value and move the
flashing digit.
Press [Tare] to increase and
[Zero] to decrease the numeric
values then press [Mode/Print]
to move onto the next setting.
E3 Ld1 Then fixed value displayed Press [Mode/Print]
E4 Lro Then fixed value displayed Press [Mode/Print]
E5 Lr1 Then fixed value displayed Press [Mode/Print]