Err 19
Weight lower limit is larger than
upper limit
High limit is set first, then the low limit is
set higher than the high limit and low
limit not equal to zero.
Err 20
PLU storage/ reading is more
than 50
PLU number above max range of 50
Err ADC
Can’t find ADC chip Damaged or missing ADC chip
Load cell is damaged.
Electronics is damaged.
---OL---
Weight over range Weight over scale range
---Lo---
Below gross zero 20e Weight below scale range