7.8 USER
TECHNICAL
PARAMETERS,
F8
F8F8
F8
S
SS
SEt
EtEt
Et
The balance has a number of technical parameters the user can adjust.
Display Description Default Value
n FIL
Filter settings - 1 to 3
1 will allow the display to update at it’s fastest
1 FIL
n ZEO
The Automatic Re zero range - 1 to 7
1 is the minimum A/Z range and 7 is the maximum
4 ZEO
n STA
Stability symbol range - 1 to 6
1 is the smallest stability range and 6 is the largest.
2 STA
1) Switch on the balance then press and hold the [Unit/Clr] key during the self-
checking test of the display.
2) After a few seconds, the display will show the first function F1 UNT.
3) Keep pressing the [Air/Liquid] key until F8 SET is displayed.
4) Press the [Tare] key to view the current setting.
5) To change the settings press the [Air/Liquid] key. Press the [Tare] key to store
the setting and advance to the next parameter. When complete the display will
show the TECH parameter.
The display will now show TECH. This parameter is for qualified technicians only and
requires entry via a pass code. Press the [Unit/Clr] key to return to weighing or
press [Air/Liquid] to go to the first function again.