11.0 ERROR CODES
During the initial power-on testing or during the operation, the scale may
show an error message. The meaning of the error messages are described
below.
If an error message is shown, repeat the procedure that caused the
message, such as turning the balance on, calibration or any other functions.
If the error message is still shown then contact your dealer for further
support.
ERROR CODE DESCRIPTION POSSIBLE CAUSES
Err 4
Initial Zero is greater than the
permissible value (typically 4%
of maximum capacity) when
power is turned on or when the
[Zero] key is pressed,
Weight on the pan when turning
the scale on.
Excessive weight on the pan
when zeroing the scale.
Improper calibration of the scale.
Damaged load cell.
Damaged Electronics.
Err 5
Keyboard error. Improper operation of the scale.
Err 6
A/D count is not correct when
turning the scale on.
Platform is not installed.
Load cell may be damaged.
Electronics may be damaged.
FAIL H or FAIL L
Calibration error Improper calibration.
If the problem persists contact
your dealer or Adam Equipment
for assistance.