Overload Warning-: please do not add any item that exceeds the maximum capacity
of the scale. When the buzzer is sounded and the display shows”-----”, remove the
item to avoid damage to the load cell and scale.
3.5
REMOTE
CONTROLLER
A: Determine increments in which the weight display
B: Hold the settled values
C: Zero the scale
D: Backlight (only used on scales with LCD display)
4.0
CALIBRATION
Hold [Zero] when the scale is powered on.
The display on the scale will show SPAN”. Make sure the hook is
empty or has the slings or chains to be used in place. See Below.
Depress the [Zero] button, the display on the scale will show “Set” then
“0” and then“0000.0” after few seconds. Zero calibration is now
complete.
Enter the weight to calibrate with by using [Hold] key to increment a
digit then the [Change/lb/kg] key to move it across the screen left.
For example, the display on the IHS will display “3000.0” for a 3 ton
calibration mass.
Add the calibrated mass on the hook, it the best that the mass is close
to the capacity where possible.
When the scale is stable, press the [Zero]. The display shows ‘HOLd’.
After a few seconds, the calibration is complete.