Product Manual

C a re and Maintenance
Disinfection: We recommend soaking in disinfection or therm o-
chemical disinfection at up to 65°C in a washer sterilizer. Adhere
strictly to manufacture rs instructions re g a rding concentration
and duration. After disinfection, thoroughly rinse in sterile water
to remove sterilizer and dry with a soft cloth. Devices can be dis-
infected at temperatures up to 95°C without addition of chemical
additives. Autoclaving: Autoclaving may be used for Nasal Spec-
ula at up to 134°C max, but will reduce the life of instru m e n t s .
A
S p e c i a l
T h a n k
Yo u . . .
Thank you for choosing an ADC 2.5v diagnostic instru-
ment. We're proud of the care and quality that goes into
the manufacture of each and every diagnostic instrument
that bears our name.
Only the finest materials are used to assure you of a
timeless instrument designed for optimum performance.
You'll quickly appreciate the results for you now own one
of the finest diagnostic instruments that money can buy.
With proper care and maintenance your ADC diagnostic
instrument is sure to provide you with many years of
dependable service.
Please read the following instructions and general infor-
mation which will prove helpful in allowing you to
enjoy your ADC product.
Thank you for your patronage. It is indeed our pleasure
to serve you.
Sincerely,
American Diagnostic Corp.
©2004 AMERICAN DIAGNOSTIC CORP.
Parts and Assembly
Introduction
This instruction manual is for use with the Complete 2.5v
I n s t rument Set (Item #5215), 2.5v Diagnostic Set (Item
#5210), 2.5v Otoscope (Item #5211 or 5211V), and all acces-
sories and replacement parts.
Battery Handle
All instrument sets come equipped with a "C" cell knurled
chrome handle with rheostatic power switch (Item #5260)
Battery Replacement:
Requires 2 "C" cell batteries. Remove the knurled cap
at the base of the handle by turning counter clockwise.
Insert two fresh "C" cell batteries (not included) with
the positive polarity towards the instrument head.
Replace battery compartment cap by turning clock-
wise. (Fig. 1).
Removing and Attaching the Instrument Head:
All instrument heads feature a bayonette style mount.
To remove, push down firmly and rotate counter-
clockwise. To engage, align slots on base of instru-
ment head with tabs on bayonette. Push head and twist
clockwise until engaged. (Fig. 2).
Turning the instrument on:
To illuminate, press the white power switch and rotate
the rheostatic bezel clockwise until desried lamp inten-
sity is obtained (0 position is off, red 1 position is full
power) Slide the switch in the opposite direction to
turn off. (Fig. 3).
Handle Care and Maintenance
The handle can be cleaned with a moist cloth or alcohol wipe.
Do NOT immerse the handle in any liquid. Periodically check
the condition of the batteries, making sure there is no sign of
corrosion or oxidation. Always replace BOTH batteries. Alka-
line batteries are recommended. Remove batteries from han-
dle if instrument will not be used for an extended period of
time.
Instrument heads
Otoscope Head (provided on 5215, 5210, 5211, and 5211V)
(Item #5220). This instrument head is designed for examina-
tion of the ear canal – specifically the external acoustic duct
and tympanic membrane. The head comes complete with a 3x
removeable viewing lens, insufflator fitting, and 3 s p e c u l a
( s t a n d a rd models with 2.5mm, 3.5mm, and 4.5mm specula,
veterinary model with 4mm, 5mm, and 7mm specula).
Attaching the specula:
Push the flanged end of the desired
specula onto the instrument head
aligning the slot in the specula
flange with the rivet at the top of the
i n s t r ument head and twist clockwise
to engage. Reverse procedure to
remove. (Fig.4).
Using Welch Allyn or equivalent disposable specula:
Attach the disposable specula adaptor, Item #5211-AD in the
same manner as the reusable specula. Attach Welch Allyn or
equivalent disposable specula by pushing the flanged end of
the desired specula onto the chrome socket and twist
counter-clockwise to engage. Reverse procedure to remove
and discard after each use.
Removing Viewing lens:
To remove during an examination, rotate
lens counter-clockwise and pull. Reverse
procedure to attach. (Fig. 5).
Replacing lamp:
Remove specula. Pull black sleeve off of
lamp. Twist lamp counter-clockwise to
remove. Replace with ADC Item #5211-4.
Reverse pro c e d u re to intall new lamp. (Fig. 6).
Insufflator
The unit comes equipped with an
insufflator fitting already attached. The
insufflator bulb (Item #5122) can be
purchased separately. (Fig. 7).
Ophthalmoscope Head
(provided on set #5215 and #5210)
This instrument head (Item #5240) is
designed for examination of the fundus, and
other ocular structures. The head is equipped
with 25 lenses from -25 to +40 diopters. (Fig. 8).
Use, Care, &
Maintenance
FORQUESTIONS,
COMMENTS, or SUGGESTIONS
CALL TOLL FREE:
1-800-ADC-2670
To Register Your Pr o d u c t ,
visit us at www. a d c t o d a y. c o m
and follow the links
Otoscope / Ophthalmoscope
I n s t r ument Set
55 Commerce Drive • Hauppauge, NY 1 1 7 8 8
w w w. a d c t o d a y. c o m
P a rt# 93-5215-00 Printed in the U.S.A.
Corrective lens selection:
Unit is equipped with 25 corrective lenses
from +40 to -25 diopters (negative lenses
in red). To select, rotate the corrective
lens selection wheel at either side of the
instrument head. Selected lenses can be
viewed in the illuminated panel. (Fig. 9).
Replacing Halogen lamp:
Remove the Ophthalmoscope head by turning counter-clock-
wise. Grab end of lamp from inside base
of Ophthalmoscope head and re m o v e .
Replace with new lamp (Item #5210-4),
taking care not to touch the glass, and
align lamp metal guides on side of lamp
with grooves in lamp socket.(Fig. 10).
Cleaning: Head can be cleaned with a moist cloth or alcohol
wipe. Remove lamp first. Do NOT immerse head in any liquid.
Bent Arm Illuminator Head:
(Item #5215-TH). This instru m e n t
head is designed for examination
of the oral cavity in conjuntion
with the tongue depressor blade,
or laryngeal mirror attachments.
Supplied attachments include a
reusable tongue blade (Item
#5215-TD), adaptor for dispos-
able tongue blade (Item #5215-TO), 3mm laryngeal mirror
(Item #5215-M3), and 4mm laryngeal mirror (Item #5215-M4).
These attachments are supplied with complete set (Item
#5215) only or can be purchased separately. (Fig. 11).
Attaching any of the supplied Attachments
Align slot on attachment with pin on top surface of bent arm
illuminator and push until secure. (Fig. 11).
Replacing lamp
Remove by turning counter-clockwise. Reverse pro c e d u re to
replace (Item #5211-4).
Care and Maintenance
D i s i n f e c t i o n : W e recommend soaking in disinfection or therm o-
chemical disinfection at up to 65°C in a washer sterilizer. Adhere
strictly to manufacture rs instructions re g a rding concentration and
duration. After disinfection, thoroughly rinse in sterile water to
remove and dry with a soft cloth. Devices can be disinfected at
t e m p e r a t u r es up to 95°C without the addition of chemical additives.
Autoclaving: Autoclaving may be used for the attachments at up
to 134°C max, but will reduce the life of the instruments.
Nasal Specula Attachment: ( p rovided on Item #5215)
This instrument head attachment (Item #5215-NS) is designed for
examination of the nasal cavity. It is used in conjunction with the
Otoscope instrument head.
Attaching Nasal Specula
This instrument must be used in conjunc-
tion with the Otoscope head (Item #5220).
Attach Otoscope head to the battery handle.
Remove ear specula or adaptor. Attach
Nasal Specula by pushing the flanged end
of the desired specula onto the instrument head aligning the slot
in the specula flange with the rivet at the top of the instru m e n t
head and twist clockwise to engage. Reverse pro c e d u re to
re m o v e . (Fig. 12).
Operating Nasal Specula
To open, turn thumb screw counter-clockwise. To close reverse pro c e d u re .
Limited Warranty
American Diagnostic Corporation (ADC) warrants its products
against defects in materials and workmanship under normal
use and service as follows:
1. Warranty service extends to the original retail purchaser
only and commences with the date of delivery.
2. The optical system is warranted for one year.
What is Covered: Repair, or replacement of parts, and labor.
What is not covere d : Tr a n s p o rtation charges to and from ADC.
Damages caused by abuse, misuse, accident, or negligence.
Incidental, special, or consequential damages. Some states do
not allow the exclusion or limitation of incidental, special, or
consequential damages, so this limitation may not apply to you.
To Obtain Wa rranty Serv i c e : Send item(s) postage paid to ADC,
Attn: Repair Dept., 55 Commerce Dr., Hauppauge, NY 11788.
Please include your name and address, phone no., proof of
p u r chase, and $2.00 to cover the cost of re t u rn shipping and
h a n d l i n g .
Implied Wa r ranty: Any implied warranty shall be limited in duration
to the terms of this warranty and in no case beyond the original
selling price (except where prohibited by law). This warr a n t y
gives you specific legal rights and you may have other rights
which vary from state to state.
(Fig. 1)
(Fig. 2)
(Fig. 3)
(Fig. 8)
(Fig. 4)
(Fig. 5)
(Fig. 6)
(Fig. 7)
(Fig. 10)
(Fig. 9)
I t e m # For Use In D e s c r i p t i o n
5 1 2 2 5 2 1 1 I n s u fflator Bulb, Otoscope
5 1 8 1 5 2 1 1 Disp. Specula, 2.5mm, 1000/pkg
5 1 8 4 5 2 1 1 Disp. Specula, 4.0mm, 1000/pkg
5 2 1 5 - T O 5 2 1 5 Tongue Depressor Attachment
5 2 6 0 5210, 5211, 5211V, 5215, 5312 B a t t e ry Handle, "C" Cell
5 2 2 0 5210, 5211, 5211V, 5 2 1 5 Otoscope Head for Otoscope, Standard,
and Complete Diagnostic Sets
5 2 4 0 5210, 5215 Ophthalmoscope Head for Standard and
Complete Diagnostic Sets
5 2 1 5 - T H 5 2 1 5 Bent Arm Illuminator
5 2 1 5 - M 3 5 2 1 5 L a ryngeal Mirro r, 3mm
5 2 1 5 - M 4 5 2 1 5 L a ryngeal Mirro r, 4mm
5 2 1 5 - N S 5 2 1 5 Nasal Speculum For Diagnostic Set
5 2 1 5 - T D 5 2 1 5 Metal Tongue Depre s s o r
5 2 1 1 - A D 5 2 1 1 Ear Specula Adaptor for Welch Allyn
5 2 1 1 - 1 5 2 1 1 Ear Specula, 2.5 mm
5 2 1 1 - 2 5 2 1 1 Ear Specula, 3.5 mm
5 2 1 1 - 3 5 2 1 1 Ear Specula, 4.5 mm
5 2 1 1 - 7 5 2 1 1 V Ear Specula, 4.0 mm
5 2 1 1 - 8 5 2 1 1 V Ear Specula, 5.0 mm
5 2 1 1 - 9 5 2 1 1 V Ear Specula, 7.0 mm
5 2 1 0 - 4 5 2 1 0 Lamp, 2.5v for Ophthalmoscope
5 2 1 1 - 4 5 2 1 1 , 5 2 1 1 V Lamp, 2.5v for Otoscope
5 2 1 0 - Z C 5 2 1 0 Zipper Case for Diagnostic Set
5 2 1 5 - Z C 5 2 1 5 Zipper Case for Complete Set
(Fig. 11)
(Fig. 12)