User`s manual
10. Certifications
Device standard: 
This device is manufactured to meet the European and United States standards for non-invasive
blood pressure monitors:
 EN1060-1 / 1995 • EN1060-3 / 1997 • EN1060-4 / 2004
 AAMI/ANSI SP10
Electromagnetic compatibility:
Device fulfills the stipulations of the International standard 
 IEC60601-1-2
Clinical testing:  
 Clinical performance tests were carried out in the U.K. “Association Protocol and  British HTP
Protocol.”
The B.H.S. (British Hypertension Society) clinical protocol was used to measure the accuracy of
this product. ADC® blood pressure units using the same measurement technology are graded “AA”
for systolic/diastolic accuracy by independent investigators using the BHS protocol.  This is the high-
est grading available for blood pressure monitors. Please see bhsoc.org for more information. (Uses
the same algorithm as ADC® B.H.S. graded “AA” model number BP3BT0-A). US patents: 6,485,429
(Irregular Heartbeat Feature) - 6,447,457 (ADC® Averaging Mode)
11. Technical Specifications
Weight: 471.5 g (with batteries)
Size: 160 (W) x 140 (L) x 98 (H) mm
Storage temperature: –20 to +50°C (-4°F - 122°F)
Humidity: 15 to 90% relative humidity maximum
Operation temperature: 10 to 40°C ( 50°F - 104°F)
Display:  LCD-Display (Liquid Crystal Display)
Measuring method: Oscillometric
Pressure sensor: Capacitive
Measuring range: SYS/DIA: 30 to 280 mmHg
Pulse: 40 to 200 per minute
Cuff pressure display range: 0–299 mmHg
Memory: Automatically stores the last 99 measurements for 2 users (total 198)
Measuring resolution: 1 mmHg
Accuracy: Pressure within ± 3 mmHg / pulse ± 5 % of the reading
Power source: a) 4 AA batteries, 1.5 V
 b) AC adapter 6 V DC 600 mA
 (voltage 4.5 V DC to 6 V DC)
Accessories: Wide range rigid cuff 8.7” – 18.1” (22 - 46 cm)
Technical alterations reserved!