Product Manual

IB p/n 9360N-00 Printed in the U.S.A.
Low Level Disinfection
Prepare Enzol enzymatic detergent according to the manufacturer’s instructions. Spray
detergent solution liberally onto cuff and use a sterile brush to agitate the detergent so-
lution over entire cuff surface for five minutes. Rinse continuously with distilled water
for five minutes. To disinfect, first follow the cleaning steps above, then spray cuff with
10% bleach solution until saturated, agitate with a sterile brush over entire cuff surface
for five minutes. Rinse continuously with distilled water for five minutes. Wipe off excess
water with sterile cloth and allow cuff to air dry.
CAUTION: Do not iron cuff.
CAUTION: Do not heat or steam sterilize cuff.
Manometer Quality Control
A Serial number and Lot number are
automatically assigned to every aneroid
during manufacturing, ensuring every item
is "controlled".
The Serial Number can be located on the
faceplate of each aneroid (Figure 4).
The Lot number is located on the the box
end label (Figure 5).
Standards
ANSI/AAMI/ISO 81060-1:2007 • ANSI/AAMI SP-10:2002
EN 1060-1:1995 • EN 1060-2:1995
Disposal
When your sphygmomanometer has reached its end of life, please be sure to dispose
of it in accordance with all regional and national environmental regulations. Devices
that have become contaminated should be disposed of in accordance with all local
ordinances and regulations.
Symbol Definitions
The following symbols are associated with your aneroid sphygmomanometer.
!
!
Warranty
The manufacturer warrants its products against defects in materials and workmanship
under normal use and service as follows:
Warranty service extends to the original retail purchaser only and commences
with the date of delivery.
Warranty duration is as follows:
Manometer Inflation System
5 YEARS 1 YEAR
10 YEARS 1 YEAR
20 YEARS 3 YEARS
LIFE 3 YEARS
*Refer to box end panel for specific warranty duration.
What is Covered: Calibration, repair, or replacement of parts and labor.
What is Not Covered: Transportation charges. Damages caused by abuse,
misuse, accident, or negligence. Incidental, special, or consequential damages.
Some states do not allow the exclusion or limitation of incidental, special, or
consequential damages, so this limitation may not apply to you.
Implied Warranty: Any implied warranty shall be limited in duration to the terms
of this warranty and in no case beyond the original selling price (except where pro-
hibited by law). This warranty gives you specific legal rights and you may have other
rights which vary from state to state.
To Obtain Warranty Service: Send item(s) postage paid to: Warranty Service
Center, 55 Commerce Dr., Hauppauge, NY 11788. Please include your name and
address, phone no., proof of purchase, and a brief note explaining the problem.
***LATEX-FREE***
LOT 0 00000 000
DIAGNOSTIX
TM
Aneroid
SIZE: Adult
COLOR: Black
CONTENTS: 1 ea
Hauppauge, NY
0000-00
ADC
0000
!
REF
(Figure 5)
Lot Number
Symbol Definition
Important Warning/Caution
Latex-Free
Circumference Size
Symbol Definition
Conforms to EU Standards
Authorized European
Represenative’s Information
Manufacturer’s Information
(Figure 4)
Serial Number
Aneroid Sphygmomanometer
Use, Care, & Maintenance