Non-Contact Thermometer Model 429 User’s Manual PLEASE NOTE: THIS MEDICAL INSTRUMENT MUST BE USED ACCORDING TO INSTRUCTIONS TO ENSURE ACCURATE READINGS.
A Special Thank You... Thank you for choosing an ADC® AdtempTM Non-Contact thermometer. We're proud of the care and quality that goes into the manufacture of each and every item that bears our name. Only the finest materials are used to assure you of a timeless instrument designed for optimum performance. You'll quickly appreciate the results, for you now own one of the finest Non-Contact thermometers that money can buy.
AdtempTM 429 - Instruction Manual Table of Contents 1. Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5 2. Introduction / Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . .5 3. General Warnings and Cautions . . . . . . . . . . . . . . . . . . . . .5-7 4. Using the Thermometer . . . . . . . . . . . . . . . . . . . . . . . . . . .8-9 4a. Controls and Features . . . . . . . . . . . . . . . . . . . . . . . . . . . .8 4b. LED Screen/Display . . . . . . . . .
1. Symbols Documentation Symbols t^okfkdW The warning statements in this manual identify conditions or practices that could lead to illness, injury, or death. `^rqflkW The caution statements in this manual identify conditions or practices that could result in damage to the equipment or other property, or loss of data. Consult instructions for use (IFU). Meets essential requirements of European Medical Device Directive 93/42/EEC.
Miscellaneous Symbols Type BF applied part Battery Level Indications Low battery (thermometer can still be used) Low battery- the batteries must be replaced before further operation 2. Introduction / Intended Use Intended use The Adtemp™ Non Contact thermometer is a clinical-grade device intended for the intermittent measurement of human body temperature in patients of all ages. It can also be used to measure the temperature of objects. 3.
Warning: If the accuracy of any measurement is in question or the use of this thermometer is not appropriate due to patient condition, check the patient's temperature with an alternate method and then check to verify the device is functioning properly by comparing it against a calibrated reference device. Warning: Keep this thermometer and its batteries out of the reach of children. Do not allow children to take their temperature with this thermometer unattended.
Caution: Do not expose the thermometer to extreme temperatures or humidity levels. Do not expose to direct sunlight. Caution: The thermometer is not waterproof; do not immerse in water or any other liquid. Caution: Avoid dropping the device or exposing it to heavy shock or vibration. Caution: Do not autoclave. Follow only the cleaning procedures described in this manual. Caution: Stop using this device if it operates erratically or if the display malfunctions.
4. Using the Thermometer 4a. Controls and Features 4b. LED Screen/Display 4c.Selecting a Scan Mode With the power on, press the Human/Object button to switch between scan modes. There are 4 modes available, including Human, Object, Human/Night, and Object/Night mode (in order). The thermometer’s volume will be muted in both Human/Night mode as well as Object/Night mode, and the Moon symbol will appear on the LCD in both Night modes.
Note: Each press of the Human/Object button is accompanied by a beep sound to confirm that the setting is activated. (excluding Night modes) 4d.Changing the Temperature Scale The Adtemp™ 429 Non Contact Thermometer can display readings in either Fahrenheit (F) or Celsius (C). The default temperature scale is Fahrenheit. To change the scale to Celsius, power the device on, then press and hold the Human/Object and Memory buttons simultaneously for 1 second.
Note: If the patient’s skin is covered with hair, sweat, or dirt, clean the area and wait 10 minutes before taking a measurement. Note: Ensure that the thermometer is held firmly during measurement and that the patient does not move until the measurement is complete. Patient movement can impact the measurement. 4. Press and release the Measure button making sure not to obstruct your view of the LED display or the buttons area. 5. Slowly move the device towards the patient’s forehead.
6.Memory Function The unit retains the last 30 readings. To recall previous readings: 1. Press the Measure button to power the unit on. 2. Briefly press and release the Memory button then press again to display the last measurement accompanied by the symbol. Note: The Human or Object symbols will be displayed to indicate the measurement mode that was in use for the displayed reading in memory. 3. Each press of the Memory button will display a previous measurement (up to 30).
7. Troubleshooting/Error Messages LCD Display Cause The temperature measured is higher than: 1. Human Mode: 122°F (50°C) 2. Object Mode: 212°F (100°C) Solution Take the measurement again. Follow the steps in Section 5. If necessary, clean the sensor lens. The temperature measured is lower than: 1. Human Mode : 50°F (10°C) 2.
If the batteries are completely depleted, “Lo” will be displayed along with the low battery symbol. 1. Use a Phillips head screwdriver to loosen battery cover screw. Remove cover. 2. Remove the batteries. 3. Insert two AAA (or equivalent) batteries, correctly positioned with positive “+” and negative “-” as shown in illustration. 4. Place the battery cover on the thermometer and tighten the screw to secure it in place. Note: Remove the batteries before storing the thermometer for an extended period of time.
Note that ADC is not responsible for validating the cleaning or disinfection protocols used on this device outside of ADC’s manufacturing site. You should determine the requirements for personnel, equipment, and materials necessary to achieve your desired disinfection levels and may need to employ routine monitoring of your disinfection protocols within your facility. After cleaning or disinfecting, wait at least 10 minutes prior to taking another measurement. 9.
11. Technical Specifications Type Non-Contact infrared thermometer Model no. 429 Measuring Range: Human Mode: Object Mode: Calibration Accuracy: Human Mode: 122.0° F 50° C 50.0°F ~ 122.0°F (10°C ~ 50°C) 6 50° F 10° C 212.0° F 100° C 32.0°F ~ 212.0°F (0°C ~ 100°C) 32.0° F 0° 7 C 71.6°F ~ 104.0°F: ±0.5°F (22°C ~ 40°C): ±0.3°C 104.0° F 40° C 71.6° F 22° C Object Mode: 68°F: ± 1°F > 68°F: ± 5% ( 20°C: ± 1°C) (> 20°C: ± 5%) Display Resolution: 0.1°F (0.1°C) Operating Environment: 60.
12. Standards ASTM 1965-98 EN 12470-5:2003 MDD (93/42/EEC) Annex II 13. Guidance and Manufacturer's Declaration The 429 IR Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the 429 Thermometer should assure that it is used in such an environment.
dìáÇ~åÅÉ=~åÇ=j~åìÑ~ÅíìêÉêDë=aÉÅä~ê~íáçå=bäÉÅíêçã~ÖåÉíáÅ=Éãáëëáçåë The 429 IR Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the 429 IR Thermometer should assure that it is used in such an environment. fããìåáíó=qÉëí Radiated RF IEC 61000-4-3 fb`=SMSM=NqÉëí=äÉîÉä `çãéäá~åÅÉ 3V/m 80MHz– 2.
oÉÅçããÉåÇÉÇ=ëÉé~ê~íáçå=Çáëí~åÅÉë=ÄÉíïÉÉå=éçêí~ÄäÉ=~åÇ=ãçÄáäÉ=oc= ÅçããìåáÅ~íáçåë=ÉèìáéãÉåí=~åÇ=íÜÉ=QOV=fo=qÜÉêãçãÉíÉê The 429 IR Thermometer is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
15.
® Distributed by: American Diagnostic Corporation 55 Commerce Drive, Hauppauge, New York 11788 Inspected in the U.S.A. Made in China tel: 631-273-9600, 1-800-232-2670 fax: 631-273-9659 AViTA Corporation 9F, No.78, Sec.1, Kwang-Fu Rd., San-Chung District, 24158 New Taipei City, Taiwan EC REP Wellkang Ltd.