User manual

12. Standards
ASTM 1965-98
EN 12470-5:2003
MDD (93/42/EEC) Annex II
13. Guidance and Manufacturer's Declaration
The 429 IR Thermometer is intended for use in the electromagnetic environment specified
below. The customer or the user of the 429 Thermometer should assure that it is used in such
an environment.
bãáëëáçåë=qÉëí `çãéäá~åÅÉ bäÉÅíêçåáÅ=ÉåîáêçåãÉåí=ÖìáÇ~åÅÉ
RF emissions CISPR 11 Group 1
RF emissions CISPR 11 Class B
Harmonic Emissions
IEC 61000-3-2
N/A
Voltage
Fluxuations/flicker
emissions
IEC 61000-3-3
N/A
The 429 IR Thermometer uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
The 429 IR Thermometer is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
The 429 IR Thermometer is intended for use in the electromagnetic environment specified
below. The customer or the user of the 429 IR Thermometer should assure that it is used in
such an environment.
fããìåáíó=qÉëí fb`=SMSMN=qÉëíÉîÉä `çãéäá~åÅÉ ÉÅíêçåáÅåîáêçåãÉåíìáÇ~åÅÉ
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Power Frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
3A/m 3A/m
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Power frequency magnetic fields
should be at le
vels characteristic of
a typical location in a typical
commercial or hospital
environment.
d
ìáÇ~åÅÉ=~åÇ=j~åìÑ~ÅíìêÉêDë=aÉÅä~ê~íáçå=bäÉÅíêçã~ÖåÉíáÅ=Éãáëëáçåë