Specifications

6-72 Error Byte Definitions
SRS Residual Gas Analyzer
Bit Description Error Code
7 No filament detected. FL7
6 Unable to set the requested emission current. FL6
5 Vacuum Chamber pressure too high. FL5
4 Not used
3 Not used
2 Not used
1 Not used
0 Single filament operation. FL0
FIL_ERR Error Byte: Filament Error Byte.
Bit Description
7 Not used
6 Parameter conflict
5 Jumper protection violation
4 Transmit buffer overwrite
3 OVERWRITE in receiving
2 Command-too-long.
1 Bad Parameter received
0 Bad command received
RS232_ERR Error Byte: Communications Error Byte