Specifications
Front Panel 4-3
SRS Residual Gas Analyzer
Front Panel
The ECU mounts directly on the probe’s feedthru flange. Its front panel is designed to
rest flat against the back surface of the probe’s flange, and it is not visible while the ECU
is locked in place.
WARNING
High Voltage inside
this unit. See manual for
safety notice.
WARNING
High Voltage inside
this unit. See manual for
safety notice.
Clearance
holes
(6 places)
Internal connectors
Locking screws
(2 places)
Probe alignment holes
Figure 2 ECU Front Panel
The front panel of the ECU box has nine holes of three different sizes, and two locking
screws.
The big hole in the center provides access to internal connectors that line up with the
feedthru connectors of the probe. The six equally-spaced holes, surrounding the big hole,
provide clearance for the six bolt heads on the back of the probe’s feedthru flange.
The two remaining holes (1/4” diameter) line up with the alignment rods of the probe and
assure the correct alignment between the ECU and the Probe during installation.
The locking screws line up with two threaded holes on the back of the feedthru flange,
and are used to lock the ECU box in place.