Whenapressuremoduleisconnected,thepressuremodule
iconwillappearinthestatusbarandbeepatthesametime.
Figure8.8showsthepressuremeasurementconnections.
Proceedasfollowstomeasurepressure:
1. If measurement is active, press once, otherwise
press twice.
2. Press the navigation keys Up/Down to select
the first
item from thepopup list, then press Enter or Done to
changethemeasurement.
3. YoucanpressUnittopullupthepressureunitlistwhich
issupportedbythecalibrator.
4. You can press Module to view the pressure module’s
information
5. To improve measurementprecision, you
can press Zero
toclearthemeasuredvalue.
8.2.9 Thermocouple Measurement (Temperature)
Thecalibratorsupportsthirteenstandardthermocouples;thermocoupletypesareidentifiedbycharacter:S,R,B,K,N,
E,J,T, C,D,G,LandU.Table53showsthetemperaturerangesandaccuraciesofthesethermocouples.Thermocouple
measurementacceptstworeferencejunctioncompensationmethods:internalreferencemodeorexternalreference
mode.
When using the in ternal reference mode, the calibrator will automatically measure the temperatureof cold
Figure8.8PressureMeasurement