Gauge Pressure
[1]
P/N
Pressure Range
Media
[2]
Accuracy(%FS)
Burst
Pressure
(psi) (bar)
V15 -15 -1.0 G 0.025 (0.05)
GP2 2 0.16 G 0.05
GP5 5 0.35 G 0.025 (0.05)
GP10 10 0.7 G 0.025 (0.05)
GP15 15 1.0 G, L
[3]
0.025 (0.05)
GP30 30 2.0 G, L
[3]
0.025 (0.05)
GP50 50 3.5 G, L 0.025 (0.05)
GP100 100 7.0 G, L 0.025 (0.05)
GP300 300 20 G, L 0.025 (0.05)
GP500 500 35 G, L 0.025 (0.05)
GP600 600 40 G, L 0.025 (0.05)
GP1K 1,000 70 G, L 0.025 (0.05)
GP2K 2,000 140 G, L 0.025 (0.05)
GP3K 3,000 200 G, L 0.025 (0.05)
GP5K 5,000 350 G, L 0.025 (0.05)
GP10K 10,000 700 G, L 0.025 (0.05)
GP15K 15,000 1,000 G, L 0.05(0.1)
GP25K 25,000 1,600 G, L 0.1 1.5×
GP30K 30,000 2,000 G, L 0.1 1.5×
GP36K 36,000 2,500 G, L 0.1 1.5×
GP40K 40,000 2,800 G, L 0.1 1.1×
Absolute Pressure
P/N
Pressure Range
Media
[2]
Accuracy(%FS)
Burst
Pressure
(psi) (bar)
AP5 5 0.35 G 0.1
AP10 10 0.7 G 0.1
AP15 15 1.0 G 0.1
AP30 30 2.0 G 0.1
AP50 50 3.5 G 0.1
AP100 100 7.0 G , L 0.1
AP300 300 20 G , L 0.1
AP500 500 35 G , L 0.1
AP1K 1,000 70 G , L 0.1
AP3K 3,000 200 G , L 0.1
AP5K 5,000 350 G , L 0.1
Differential Pressure
P/N
Pressure Range
Media
Accuracy
(%FS)
Burst
Pressure
Static
Pressure
Range
(inH
2
O) (mbar)
DP1 ±1 ±2.5 G 0.05 100× ±10 psi
DP2 ±2 ±5.0 G 0.05 100× ±10 psi
DP5 ±5 ±10 G 0.05 50× ±10 psi
DP10 ±10 ±25 G 0.05 20× ±10 psi
DP20 ±20 ±50 G 0.05 20× ±10 psi
DP30 ±30 ±75 G 0.05 20× ±10 psi
DP50 ±50 ±160 G 0.05 ±10 psi
DP150 ±150 ±350 G 0.025 50 psi
DP300 ±300 ±700 G 0.025 50 psi
Electrical Measurement Accuracy
Range Accuracy
Voltage DC ±30.0000V ± (0.01%RD + 1.5 mV)
Current DC ±30.0000mA ± (0.01%RD + 1.5 μA)
DC 24V 24V±0.5V max:50mA Protect at: 120mA
Switch
[1]
Status OPEN/CLOSED
Compound Pressure
P/N
Pressure Range
Media Accuracy(%FS)
Burst
Pressure
(psi) (bar)
CP2 ±2 ±0.16 G 0.05
CP5 ±5 ±0.35 G 0.025 (0.05)
CP10 ±10 ±0.7 G 0.025 (0.05)
CP15 ±15 ±1 G 0.025 (0.05)
CP30 -15 to 30 -1 to 2 G 0.025 (0.05)
CP100 -15 to 100 -1 to 7 G, L 0.025 (0.05)
CP300 -15 to 300 -1 to 20 G, L 0.025 (0.05)
Pressure Ranges
Note: [1]. Sealed gauge pressure for above 1000 psi
[2]. G=Gas, L=Liquid
[3]. 0.025% FS for gas media only
Note: [1] 1V~12V if switch has detective voltage
SPECIFICATIONS
* One year accuracy (including yearly stability) except DP1 and DP2 range
which is 0.05%FS calibration accuracy and 0.05%FS yearly stability.