Open the reservoir
The media level should be 2/3 the height
of the reservoir.
B
Filling/emptying the reservoir
Clean any impurities from the media and reservoir.
C
Lock the reservoir
Preparation
Basic Operation
Connection
A
Vent
B
Zeroing should
be operated
under this
condition only.
open
open
DIGITAL PRESSURE CALIBRATOR
Units
A
0255075100
P
ressu
re
G
au
ge
Pre-pressure
E
DIGITAL PRESSURE CALIBRATOR
Units
A
0255075100
P
ressu
re
G
au
ge
open
open
④turn it to the
mid position
③turn it to
greatest
degree
①open
②open the upper cover
Fill or empty media
①close the upper cover
②close
Vacuum preparation
G
0
0.02
0.04
0.06
0.08
-0.1
Vacuum
G
auge
DIGITAL PRESSURE CALIBRATOR
Units
A
0 25 50 75 100
②open
①open
④turn it to the
mid position
③turn it to the
smallest
degree
Vacuum process
H
0
0.02
0.04
0.06
0.08
-0.1
Vacuum
G
auge
DIGITAL PRESSURE CALIBRATOR
Units
A
0255075100
③fine adjust
①close
②turn it to approach
the desire value
SW V mA COM 24VSW V mA COM 24V
SW V mA COM 24V
dial gauge
pressure switch pressure transmitter
k
Pressure Gauge
DIGITAL PRESSURE CALIBRATOR
Units
A
0255075100
flexible hose
Pressurizing process
F
DIGITAL PRESSURE CALIBRATOR
Units
A
0255075100
P
ressu
re
G
a
u
ge
①close①close
If necessary,
please generate
high pressure
using #4 handle
③generate high-pressure
and fine adjust pressure
②Pre-pressurization
Purge process
C
DIGITAL PRESSURE CALIBRATOR
Units
A
0255075100
①close
③turn it to the smallest degree
②Turn all the
way out
Purge process (cont.)
D
DIGITAL PRESSURE CALIBRATOR
Units
A
0255075100
②open
①open
If air is still trapped in the pump
then repeat the purge process 2
or 3 more times until all the gas
is removed from the system.
,
③Turn all the way in
bubble
Additel Corporation.USA
www.additel.com
If it is difficult to generate
vacuum pressure then
repeat the purging
process to make sure all
the gas is removed from
the system.
,
Vacuum displayed on the gauge
Wait for gauge to vent to zero