Scalar 1000 Library Operator Guide
Copyright Notice © Copyright ADIC 2000 The information contained in this document is subject to change without notice. This document contains proprietary information which is protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without prior written consent of ADIC.
Contents 1 Introduction Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3 Intended Audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3 Organization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3 Associated Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Functional Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6 Tape Drives . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6 Control Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7 Cartridge Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7 Cartridge Accessor .
Scalar 1000 Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-37 Scalar 1000 Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-38 Normal Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-38 Emergency Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Insert/Eject Submenu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-54 Insert Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-55 Insert Clean Tape Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-56 Eject Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-61 Eject Clean Tape Dialog . .
Initialize Submenu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-127 Communication. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-128 SCSI. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-129 Service Menu Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-130 Start Dialog . . . . . . . . . .
Direct Insertion into the Control or Expansion Module . . . . . . . . . . . . . . . . . . . . . 6-4 Ejecting Media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4 Insert/Eject Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5 Direct Removal from the Control or Expansion Module . . . . . . . . . . . . . . . . . . . . 6-6 Processing Media . . . . . . . . . . . . . . . . . .
Figures Figure 2-1 Scalar 1000 Frames . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4 Figure 2-2 Coordinate System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9 Figure 2-3 Half Inch Storage Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10 Figure 2-4 8mm Storage Cell. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11 Figure 2-5 Gripper . . . . . . .
x Figures Figure 5-1 Directory Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6 Figure 5-2 Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8 Figure 5-3 Mode Dialog. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9 Figure 5-4 Ready Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
01 Aug 2000 Figure 5-34 Response Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-33 Figure 5-35 Response Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-34 Figure 5-36 Response Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-34 Figure 5-37 Error Log Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
xii Figures Figure 5-67 Insert Range Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-57 Figure 5-68 Target Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-58 Figure 5-69 Usage Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-59 Figure 5-70 Response Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
01 Aug 2000 Figure 5-100 Password Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-86 Figure 5-101 Security is OFF Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-87 Figure 5-102 Change Password Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-88 Figure 5-103 Password Changed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-89 Figure 5-104 Media Dialog . .
xiv Figures Figure 5-133 Clean Drives Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-123 Figure 5-134 Progress Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-124 Figure 5-135 Response Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-124 Figure 5-136 Unload Drives Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
01 Aug 2000 Figure 5-166 Response Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-151 Figure 5-167 Step Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-152 Figure 5-168 Continuation Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-152 Figure 5-169 Scanner Submenu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6-1 xvi Figures Barcode Label Examples. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tables Table 2-3 Element Coordinates for Firmware 2.3 or higher . . . . . . . . . . . . . . . 2-8 Table 2-4 Library Element Indexing for Firmware 2.30 or Below . . . . . . . . . 2-12 Table 2-5 Library Element Indexing for Firmware 3.0 or Higher . . . . . . . . . 2-12 Table 2-6 Scalar 1000 Configurations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16 Table 3-1 Hazard Alert Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
xviii Tables 600561-A
1 Introduction Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3 Intended Audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3 Organization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3 Associated Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-2 Introduction 600561-A
1- Overview This manual contains information and instructions necessary for the safe operation of the Scalar 1000 1 library. The topics discussed in this chapter are: • Overview • Intended Audience • Organization • Associated Documents • Explanation of Symbols and Notes • Assistance Intended Audience This guide is intended for operators, trained customer specialists, and maintenance personnel of the service partner who interact with the Scalar 1000.
Chapter 5 Menus and Commands - Describes the visual menus and commands executed by the Scalar 1000. Chapter 6 Processing Media- Describes the Insert/Eject station, various media, and processing of the media. Chapter 7 Error Messages - Describes message processing and troubleshooting hints.
Assistance Warning Operation of the Scalar 1000 by untrained personnel can lead to dangerous situations. The consequence could be severe or fatal injury caused by contact with electrical connections. Therefore, introductory training by ADIC is recommended for all who work with the Scalar 1000.
if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. To meet FCC emissions limits, properly shielded and grounded cables and connectors must be used.
2 Description Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3 Subsystem Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3 Host Attachment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4 Remote Management Unit (RMU) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-2 Description 600561-A
2- Overview An in-depth description of the subsystem, tape cartridge, storage and drive configuration, and optional features are provided. Subsystem Description The Scalar 1000 is a library that automates the retrieval, storage and control of half-inch type cartridge tapes (8490, 3480, 3590/E), DLT (4000, 7000, and 8000) and 8mm (AIT) type cartridges. If the library is configured for 8mm cartridges, no other cartridge type may be used in the library.
Figure 2-1 Scalar 1000 Frames Host Attachment The Scalar 1000 contains two independent SCSI device adapters that offer additional flexibility during installation. The library can be configured with either the High Voltage Differential (HVD) or Low Voltage Differential/Single Ended (LVD/SE) SCSI card, depending on the host adapter. Each SCSI bus must be terminated with the appropriate HVD or LVD/SE terminator.
As an alternative, the Scalar 1000 can be network attached through an Archive Management Unit (AMU) in a main frame, UNIX, or Windows NT environment. Contact a sales representative for additional information about using the AMU. The Scalar 1000 can also be connected to a Fibre Channel Host via the ADIC Fibre Channel Router (FCR). The ADIC FCR provides one or two SCSI parallel bus connections and a Fibre Channel connection, translating (bridging) the SCSI protocol between these two media types.
Functional Description The Scalar 1000 can be installed on a solid or a raised floor. The cartridge accessor moves cartridges between storage cells, tape drives and the Insert/Eject Station (I/E Station). The cartridge accessor includes the gripper. A barcode scanner on the gripper identifies the cartridges in cells. Refer to Cartridge Accessor on page 2-12 for additional details.
• Model 8490 (half inch) • Model 8590 (half inch) • IBM 3590/E (half inch) • Plasmon NCTP (half inch) • Model 4001 (DLT) • Model 4001S (DLT) • Model 7001 (DLT) • Model 7001S (DLT) • Model 8001S (DLT) • AIT3102 (AIT 8mm) • AIT5002 (AIT 8mm) • AIT5102 (AIT 8mm) Control Hardware The library control hardware controls all operations in the Scalar 1000, including the interaction between the library and operator and the library and the host.
Table 2-3 Element Type Element Coordinates for Firmware 2.3 or higher ID Frame (Module) Section Column Row Storage S 01 - 04 1-4 A-E 01 - 18 I/E Station I 01 1 A 01 - 18 Drives D 01 - 04 1-2 A-B 01 - 06 The Element Type can be I for I/E Station, D for Drive, and S for Storage. The frame number can be from 1 to 4. The Sections are numbered, for storage, from 1 to 4 top to bottom, and for drives, 1 to 2 bottom to top.
Drive Drive Column Column A B Drive Bay 1 Drive Drive Column Column A B Drive Bay - Starts from bottom to top Figure 2-2 Column E Column D Column C Column B Column A 1x10 or 1x15 Storage Array Drive Bay 2 5x10 or 5x15 Storage Array 1x10 or 1x15 Storage Array Column E Column D Column C Column B 1x10 or 1x15 Storage Array 1x10 or 1x15 Storage Array Column A 5x10 or 5x15 Storage Array 5x12 or 5x18 Storage Array Storage Section 3 4x12 or 4x18 Storage Array Storage Section 2 Storage Secti
Figure 2-3 2-10 Description Half Inch Storage Cell 600561-A
Figure 2-4 8mm Storage Cell To manipulate the media within the library, the host must reference each movement with source and target designations. This is done via element addressing, which specifies precisely which cells within the library are to be used. Toward that end, the following addressing scheme will be used for the library. Refer to Table 2-4 on page 2-12 for libraries with firmware 2.3 or below and Table 2-5 on page 2-12 for libraries with firmware 3.0 or higher.
Table 2-4 Library Element Indexing for Firmware 2.30 or Below Indexing Range Element Type Half-Inch 8mm Storage 0 - 787 0 - 1181 Insert/Eject Station 788 - 799 1182 - 1199 Drives 800 - 847 1200 - 1247 Cartridge Accessor 848 1248 Table 2-5 Library Element Indexing for Firmware 3.
Figure 2-5 Gripper The cartridge accessor has: • A gripper assembly for mounting a cartridge gripper, and a barcode scanner. • • 01 Aug 2000 A cartridge gripper picks and places cartridges in storage cells, tape drives or the Insert/Eject station. A barcode scanner reads the external labels on the cartridges. The barcode scanner is used during the inventory and teach processes.
Insert/Eject Station The Insert/Eject Station allows insertion and ejection of cartridges without interrupting the normal operation of the library. The I/E Station has a capacity of 12 half-inch cartridges or 18 8mm cartridges. For a Mixed Media library of half-inch and DLT cartridges, the top 6 slots of the Insert/Eject Station are reserved for half-inch cartridges and the bottom 6 slots are reserved for DLT cartridges. Refer to Figure 2-6 on page 2-14.
Tape Cartridge Each tape cartridge in the Scalar 1000 must have an external label that is operator and machine readable to identify the bar code (BC) number. The external label contains a minimum of 5 characters and a maximum of 16 characters for the BC depending on the operating modes.
Cartridge Storage and Tape Drive Configurations Cartridge storage quantity varies by number of modules and tape drives in the modules. Table 2-6 lists the quantity of media contained by the storage cells for the Control Module and Expansion Modules configured for the different media types. Note The Scalar 1000 consists of a Control Module and up to 3 Expansion Modules.
Optional Features Optional features for the Scalar 1000 are as follows: 01 Aug 2000 • ADIC Fibre Channel Router (FCR) • Dual DC Power Supplies Optional Features 2-17
2-18 Description 600561-A
3 Safety Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3 Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3 Hazard Alert Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3 Area of Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-2 Safety 600561-A
3- Overview Avoid dangerous situations while operating the Scalar 1000 by following all safety procedures and operating the equipment carefully. Read and follow all instructions in this guide. Note In addition to the safety instructions in this guide, local and professional safety rules apply. It is mandatory that operators of the Scalar 1000 library understand and comply with all instructions contained in this guide and other related documentation.
Table 3-1 Symbol Hazard Alert Message Damage to ...
Note This symbol draws attention to user tips. No dangerous or damaging consequences for personnel or property are associated with this symbol. This symbol indicates important or useful information. No dangerous or damaging consequences for personnel or property are associated with this symbol. Area of Application Note Other manufacturer documentation is an integral part of the Scalar 1000 documentation set. This information applies to the entire Scalar 1000 family.
Main Power Switch Switching off the Main Power Switch removes all power and causes the movement electronics to shut off. All movements of the accessor stop immediately. In case of danger to personnel or property, immediately switch off the Main Power Switch. 3-6 Caution Except in emergencies, stop the Scalar 1000 with the normal shutdown procedure before switching off the Main Power Switch. ADIC is not responsible for damage caused by improper use of the Main Power Switch.
4 Operation Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3 Operator Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3 Remote Management Unit (RMU) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8 System Administrator Responsibilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-2 Operation 600561-A
4- Overview Normally, the host issues commands to the Scalar 1000. Operator control is provided via the Operator Panel. The Operator is responsible for: • Starting the Scalar 1000 • Shutting down the Scalar 1000 • Handling Media Refer to Inserting Media on page 6-3 and Ejecting Media on page 6-4 for media handling procedures. In the case of equipment failures, the operator can perform media processing.
4 X 20 LCD Panel Speaker Vent Figure 4-1 4-4 Description Operator Panel 600561-A
Table 4-1 provides an illustration and description of the push buttons associated with the Operator Panel. Table 4-1 Push button Operator Panel Push Buttons Description The Up Arrow push button can be used to: • scroll the display to show previous line(s). • move the cursor to an item before the current item when selecting an option, selection is represented by the character >. • increment the current value to the next value when used in an entry field.
Table 4-1 Push button Operator Panel Push Buttons (Continued) Description The Ready push button can be used to: • transition the library from a Ready state to NotReady or NotReady to the Ready state. Holding down the push buttons will cycle through the options. The longer the push button is held down, the faster it will cycle. Table 4-2 provides an illustration and description of the visual indicators on the Operator Panel.
The meaning of the special characters displayed on the Operator Panel are as follows: > When this character precedes the option, the option is available for selection. Pressing the button selects the option. Pressing the button presents informative text about the option. Pressing the or button selects between the respective previous or next option. ... When these characters follow the option, additional options or status displays are presented after the selected option.
Remote Management Unit (RMU) The RMU resides in each system and is pre-installed at the factory. (Please contact ADIC for pricing and installation on this option for all chassises manufactured prior to August 2000.) The RMU allows remote access via a Web browser to the library. The RMU supports Microsoft Internet Explorer version 4.0 and above, as well as Netscape Navigator versions 4.01, 4.5, and 4.73.
System Administrator Responsibilities The System Administrator (SA) must setup the RMU for proper operation. The SA responsibilities includes establishing a network address for the RMU and establishing the Uniform Resource Locators (URL). The network address consists of an Internet Protocol (IP) address, subnet mask, and gateway IP Address. Refer to RMU Submenu on page 5-93 for information about setting the network address via the Operator Panel.
• Operator Panel • Logs The Status Tab is an open dialog. The remaining five tabs, which are password protected. These tabs require a valid user or administrator account and a successful login. Refer to Center Navigation Frame on page 4-18 for the dialogs and hyperlink descriptions.
Startup Display Figure 4-2 01 Aug 2000 Remote Management Unit (RMU) 4-11
4-12 Description 600561-A Description This hyperlink logs out the current user and returns to the Status dialog. This hyperlink opens a new page and displays a brief description of the Status, Configuration, Firmware, Diagnostic files, Operator Panel, and Logs tabs. Refer to Figure 4-3 on page 4-13. This hyperlink opens a new page for the ADIC web site, redirects the user, and displays the library documentation file hyperlink. Refer to Figure 4-4 on page 4-14.
Remote Management Unit (RMU) Help Content : • Status The status page shows the current status of the Remote Management Unit, the attached library, and any drives within that library. From this page the user can easily see if there are any problems with the system. • Configuration (Password Protected) The configuration page allows the user to configure the Remote Management Unit. Network configuration (including SNMP), user configuration, and date/time setup are all on this page.
4-14 Description 600561-A Figure 4-4 Scalar 1000 Documentation Files Dialog
01 Aug 2000 Remote Management Unit (RMU) 4-15 Description This hyperlink opens an email dialog to ADIC technical support center for North America. This hyperlink opens an email dialog to ADIC technical support center for the remainder of the world. ADIC Technical Assistance Center atac@adic.
4-16 Description 600561-A Figure 4-6 Version Dialog
01 Aug 2000 Remote Management Unit (RMU) 4-17 Description This location indicates the URL identifier of the RMU site and the attached library type. Hyperlink Name The Top Information Frame dialog and description are indicated below.
4-18 Description 600561-A All of the Center Navigation Frame dialogs and descriptions, beginning with Status, are indicated below.
01 Aug 2000 Remote Management Unit (RMU) 4-19 This field value indicates the type and quantity of tape drives currently installed in the attached library. This field value indicates the name and location of the current user. This field value indicates the Hostname used for the RMU connection. This field value indicates the Internet Protocol (IP) address for the RMU connection. This field value indicates the MAC address of the RMU. This is also the serial number of the RMU.
Login Dialog Figure 4-7 4-20 Description 600561-A
01 Aug 2000 Remote Management Unit (RMU) 4-21 The user enters his/her user name. The user enters his/her password. The user submits the name and password for login verification. If either user name or user password are not verified, the login is reject. Enter Login Name Enter your password Submit The RMU is a direct interface to the attached library.
4-22 Description 600561-A Figure 4-8 Configuration Dialog The Configuration dialogs and descriptions are indicated below.
01 Aug 2000 Remote Management Unit (RMU) 4-23 This section allows the user to enter a new User Name. This pull-down menu allows the user to select a user name that has already been created. Only the current user name is listed unless the SA is logged on. Select One This section allows the user to perform management actions on the user selected. Edit New User Name This selection indicates that the current user will be deleted. This can only be done by the SA.
4-24 Description 600561-A The user types his password in again for verification. This field value allows the user to change the current date. This field value allows the user to change the current time. The user submits the configuration data for verification and allows the user to confirm any configuration changes. This button will clear all the changed fields.
01 Aug 2000 Remote Management Unit (RMU) 4-25 Figure 4-9 Confirm Configuration Changes Dialog
4-26 Description 600561-A This button confirms the configuration changes. Confirm When the user enables DHCP, the DHCP Enabled dialog is displayed. Refer to Figure 4-10 on page 4-27. The user must enter their password.
DHCP Enabled Dialog Figure 4-10 01 Aug 2000 Remote Management Unit (RMU) 4-27
4-28 Description 600561-A Figure 4-11 Firmware Dialog The Firmware dialog and descriptions are indicated below.
01 Aug 2000 Remote Management Unit (RMU) 4-29 The Firmware tab updates the firmware version of the RMU or the library. The fields and values that are associated with this tab are indicated below This section allows the user to select which firmware is to be upgraded. This section allows the user to select or browse to find the firmware upgrade file. When this button is clicked, the upgrade firmware process is started.
4-30 Description 600561-A Figure 4-12 Diagnostics File Dialog The Diagnostic file dialog and descriptions are indicated below.
01 Aug 2000 Remote Management Unit (RMU) 4-31 This selection allows the user to choose which log to retrieve. After selecting the log, click on this button to retrieve the log file. Retrieve selected files The Diagnostic files tab displays the library command error log, library error log, or RMU error log.
4-32 Description 600561-A Figure 4-13 Operator Panel Dialog The Operator panel dialog and descriptions are indicated below.
01 Aug 2000 Remote Management Unit (RMU) 4-33 The Previous button can be used to: Previous Enter Next Ready Escape The Operator panel tab displays the library operator and duplicates operator panel functionality. Access to this menu does not lock out the library’s front panel. The current state of the library menu is echoed and continuously refreshed on the display.
4-34 Description 600561-A Help • display help text for the current selected item if available.
01 Aug 2000 Remote Management Unit (RMU) 4-35 Figure 4-14 Logs Dialog The Logs dialog and descriptions are indicated below.
4-36 Description 600561-A Description The Logs tab displays the library logs.
Scalar 1000 Startup Use the following procedure to start the Scalar 1000. Caution Warning With the RMU installed, Power Up must be delayed by at least 10 seconds after a power down condition to allow the RMU to properly shutdown. Step 1 Ensure that all doors are closed and mechanically locked on the Control Module and any Expansion Modules. Step 2 Locate the Main Power Switch at the lower right corner in the back panel door of the Control Module.
Scalar 1000 Shutdown The Scalar 1000 may be shut down normally or by the Emergency Shutdown method. Attention Use the method described in Emergency Shutdown in cases of emergency only. Normal Shutdown To properly shut down the Scalar 1000 follow the procedure. Caution With the RMU installed, power up must be delayed by at least 10 seconds after a power down condition to allow the RMU to properly shutdown. Step 1 If necessary, use the button to return to the Main Menu.
Step 4 Repeat Step 2 and Step 3 for each additional Expansion Module. Emergency Shutdown Use the following procedure if an emergency situation arises. Attention Except in emergencies, stop the Scalar 1000 with the normal shutdown procedure before switching off the Main Power Switch. ADIC is not responsible for damage caused by improper use of the Main Power Switch. All risk lies entirely with the user.
4-40 Description 600561-A
5 Menus and Commands Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5 Using the Operator Panel Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5 Common Help Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7 Operator Intervention Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Insert/Eject Submenu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-54 Insert Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-55 Insert Clean Tape Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-56 Eject Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-61 Eject Clean Tape Dialog . .
Initialize Submenu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-127 Communication. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-128 SCSI. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-129 Service Menu Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-130 Start Dialog . . . . . . . . . .
5-4 Menus and Commands 600561-A
5- Overview The Operator Panel provides a menu driven operator interface to the Scalar 1000. The menus allow selection of parameters and information regarding the status and current operating parameters of the Scalar 1000. Using the Operator Panel Menu Note The information menus are not dynamically updated. To view changes, reselect the menu that was changed. 01 Aug 2000 Each menu is accessible through the Operator Panel push buttons.
Directory Structure Figure 5-1 5-6 Menus and Commands 600561-A
Common Help Button The Help key is common to all menus, submenus, dialogs, and screens. Using the Help key while the selection indicator (< or >) is at a desired location, provides informative text about that entry. To use the Help button, follow these steps. Step 1 If necessary, press the button or the button to position the selection indicator at the desired location. Step 2 Press the button to view the help text. The help text appears. Step 3 Press the button.
Main Menu Selection Note The button can always be used to escape to the previous menu. The Main Menu is the first available menu after the Initial Screen displays. Refer to Scalar 1000 Startup on page 4-37 for information about the Initial Screen. To display the Main Menu from the Initial Screen or any other menu, press the button until the Main Menu appears. The Main Menu defaults to the Mode selection. All subsequent menus and options are selectable from the Main Menu. Eight options are available.
Mode Dialog Selection The Mode Dialog is primarily used to change the operating mode of the Scalar 1000 for startup, shutdown, or diagnostic operations. Mode is selected from the Main Menu by following this procedure. Step 1 If necessary, press the Menu appears. button until the Main The Main Menu appears with the selection indicator positioned at Mode. Step 2 Press the button. The Mode Dialog appears. See Figure 5-3. The first line of the Mode Dialog displays the current state.
Press the Step 4 button to confirm the change. Repeat Step 3 for the desired parameters in the list. When toggling between the Ready and NotReady states, the following two screens, refer to Figure 5-4 on page 5-10 and Figure 5-5 on page 5-10, appear. When Shutdown is selected, the Shutdown Screen appears. See Figure 5-6.
Table 5-1 State Ready Not Ready Shutdown 01 Aug 2000 Operating State and Mode Mode Description Online The normal host controlling condition. Host commands are processed. Offline The normal operator controlling condition. Operator commands are processed. Most host commands are not processed. Online Aisle power is not present. Host commands not involving the accessor are processed. Offline Aisle power is not present. Most operator and host commands are not processed.
Status Menu Selection The Status Menu is primarily used to provide various operating statistics. Status is selected from the Main Menu by following this procedure. Step 1 If necessary, press the Menu appears. button until the Main The Main Menu appears with the indicator positioned at Mode. Step 2 Press the button or the button to position the selection indicator at Status. Step 3 Press the button. The Status Menu appears. This menu offers two selections. See Figure 5-7.
Library Submenu When Library is selected, the Library Submenu is displayed. This submenu offers four selections. See Figure 5-8. The Library Submenu defaults to the SCSI selection. >SCSI System... Element... Logs Figure 5-8 Library Submenu To choose one of the four selections, follow the steps. Step 1 Press the button to select the SCSI Submenu. . — or — Press the button or the button to select the System Dialog, Element Dialog, or Logs Submenu. Press the button. The selected screen appears.
SCSI Submenu When SCSI is selected, the SCSI Submenu is displayed. This submenu offers two selections. See Figure 5-9. The SCSI Submenu defaults to the Mode Parms selection. >Mode Parms... Reservations... Figure 5-9 SCSI Submenu To choose one of the two selections, follow the steps. Step 1 Press the Dialog. button to select the Mode Parms — or — Press the button or the Reservations Dialog. Press the button to select the button. The selected menu appears.
Mode Parms Dialog When Mode Parms is selected, the Mode Parms Dialog is displayed. See Figure 5-10. This option is used to view the changeable Mode Parameter pages. Refer to the document number 600974 Scalar 1000 SCSI Reference Manual for additional information about the Scalar 1000 Mode Parameters. Select Mode Parms: ELEMENT ADDRESS < and Type: CURRENT Accept: N Figure 5-10 Mode Parms Dialog Step 1 Press the button to select the current parameter value and move to the next parameter.
Element Address When Element Address is selected, the Element Address screen appears. Figure 5-11 shows the Element Address/ Current Screen. Figure 5-12 shows the Element Address/ Default Screen. Figure 5-13 on page 5-17 shows the Element Address/Saved Screen.
Base DT Base Base Base Base Figure 5-13 MT ST IE DT : : : : Default base address of the tape drive(s) 00001 04096 00016 00256 Element Address/Saved Screen Parameters Values Base MT Saved base address of the accessor Base ST Saved base address of the Storage Cells Base IE Saved base address of the Insert/Eject station Base DT Saved base address of the tape drive(s) Mixed Media When Mixed Media is selected, the Mixed Media screen appears. Figure 5-14 shows the Mixed Media/Current Screen.
mixed media mode. While the library is not operating in mixed media mode, the Extended RES, Extended ASCQ, and Extended VolTag are not applicable (N/A).
Extended RES Yes indicates that Extended Read Element Status defaults to enabled No indicates that Extended Read Element Status defaults to disabled Extended ASCQ Yes indicates that Vendor Unique Additional Sense Code Qualifiers default to enabled No indicates that Vendor Unique Additional Sense Qualifiers default to disabled Extended VolTag Yes indicates that the reported bar code will include media type identifier No indicates that the reported bar code will not include the media type identifier Mix
Extended ASCQ Yes indicates that Vendor Unique Additional Sense Code Qualifiers are saved as enabled No indicates that Additional Sense Qualifiers are saved as disabled Extended VolTag Yes indicates that the reported bar code will include media type identifier No indicates that the reported bar code will not include the media type identifier Parity When Parity is selected, the Parity screen appears. Figure 5-17 shows the Parity/Current Screen. Figure 5-18 on page 5-21 shows the Parity/ Default Screen.
Parity : Yes Retries: 001 Figure 5-18 Parity/Default Screen Parameters Values Parity Yes indicates that SCSI parity defaults to enabled No indicates that SCSI parity defaults to disabled Retries Indicates the default number of transmission retries (000 ..
LCD When LCD is selected, the LCD screen appears. Figure 5-20 shows the LCD/Current Screen. Figure 5-21 shows the LCD/ Default Screen. Figure 5-22 on page 5-23 shows the LCD/ Saved Screen. Security: Disabled Line : 1< MAIN SCREEN (current) Figure 5-20 Note Press the LCD/Current Dialog Parameters Values Security Disabled indicates that LCD security is currently disabled Enabled indicates that LCD security is currently enabled Line Indicates the currently displayed text line (1 ..
Text Corresponding line of main screen text with up to 20 characters Security: Disabled Line : 1< (saved) Figure 5-22 Note Press the LCD/Saved Dialog Parameters Values Security Disabled indicates that LCD security is saved as disabled Enabled indicates that LCD security is saved as enabled Line Indicates the saved displayed text line (1 .. 4) Text Corresponding line of main screen text with up to 20 characters button or the button to toggle between the line number parameter values.
Reservations Dialog When Reservations is selected, the Reservations Dialog is displayed. See Figure 5-23. This option is used to determine the state of the element reservations made by a SCSI host, to reserve elements for exclusive use within a multi initiator environment. Bus : 0 < Host ID : 0 Reservation ID: Any Found : 0000 Accept: N Figure 5-23 Reservation Dialog Step 1 Press the button to select the current parameter value and move to the next parameter.
See Figure 5-24. Otherwise, the changed parameters continue to display but no action is taken. Bus: 0 < Host ID : 0 ReservationID: Any Found : 0109 Accept : N Figure 5-24 01 Aug 2000 Response Dialog Parameters Values Found Indicates the number of reserved elements which match the Bus, Host ID, and Reservation ID parameters.
System Dialog When System is selected, the System Dialog is displayed. See Figure 5-25. This option is used to view the current library operating statistics. Cartridges Free cells Total gets Figure 5-25 : 0122 :0036 :00000137 [more] System Dialog Step 1 Parameters Values Cartridges Indicates the number of cartridges in the storage cells (000 .. 0787 for half inch and DLT cartridges) (0000 ..
Step 2 Parameters Values Moves Total move media commands that have been executed X meters Total meters traveled by the accessor in the horizontal direction Y meters Total meters traveled by the accessor in the vertical direction [more] More information on the Continuation Screen Press the Screen. button to select the Continuation The Continuation Screen appears. See Figure 5-27.
Elements Dialog When Elements is selected, the Elements Dialog is displayed. See Figure 5-28. This option queries the individual element statistics. Individual element information is tracked since the last Teach New operation. Enter Desired Coord: S< 01 2 A 01 OR Element : 00000 Accept: N Figure 5-28 Element Dialog Step 1 Press the button to select the current parameter value and move to the next parameter.
Element Indicates the element number which corresponds with the coordinate parameter Accept Y to accept changes N to reject changes Press the Step 2 button to confirm the changes. Repeat Step 1 for all of the parameters in the list. If Y is selected for the Accept parameter value, the Response Dialog appears. See Figure 5-29. Otherwise, the changed parameters continue to display but no action is taken.
Step 3 Press the Screen. button to select the Continuation The Continuation Screen appears. Refer to Figure 5-30 on page 5-30. Total Puts : 0000011 Put Retries: 0000000 Get Retries: 0000000 [more] Figure 5-30 Continuation Screen Step 4 Parameters Values Total Puts Indicates the total number of puts (0000000 .. 99999999) that occurred at this element Put Retries Indicates the total number of put retries (0000000 ..
01 Aug 2000 Parameters Values Reserved Yes indicates the element is reserved by a SCSI host No indicates the element is not reserved. When not reserved, Reserve ID and Reserved by parameters are not applicable (N/A).
Logs Submenu When Logs is selected, the Logs Submenu is displayed. This submenu offers two selections. See Figure 5-32. The Logs Submenu defaults to the Command Log selection. >Command Log... Error Log... Drive Error Log... Figure 5-32 Logs Submenu To choose one of the three selections, follow the steps. Step 1 Press the Dialog. button to select the Command Log — or — Press the button or the button to select the Error Log Dialog or Drive Error Log Dialog. Press the button. The selected menu appears.
Command Log Dialog When Command Log is selected, the Command Log Dialog is displayed. See Figure 5-33. This option is used to view a running history log of library operations. A service representative can retrieve this data for problem analysis. 11/10/99 18:25:24.88 SCSIDONE INQ : Status good Figure 5-33 Command Log Dialog To choose one of the selections, follow the steps. Step 1 Press the button or the through the entire log.
Parameters Values Continue Y to print the log to the serial port N to not print the log to a serial port Press the button to confirm the changes. If Y is selected for the Continue parameter value, the two Response Screens appears. See Figure 5-35 and Figure 5-36. Otherwise, the changed parameter continue to display but no action is taken. Note These screens may flash by quickly. Requested command is now in progress... Figure 5-35 Response Screen Requested command is now complete...
Error Log Dialog When Error Log is selected, the Error Log Dialog is displayed. See Figure 5-37. This option is used to view a running history log of error conditions that have occurred. 02/19/97 18:25:24.88 Type : SW Temp Error : 0X02040000 Modifier: 0X00000000 Figure 5-37 Error Log Dialog Step 1 Press the button or the button to scroll through the Error Log parameters. Parameters Values Type Indicates the type of error. Record the error type in Error Log Codes on page 7-15.
Step 2 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list. Parameters Values Continue Y to print the log to the serial port N to not print the log to a serial port Press the button to confirm the changes. If Y is selected for the Continue parameter value, the two Response Screens appears. See Figure 5-35 and Figure 5-36.
Drive Error Log Dialog When Drive Error Log is selected, the Drive Error Log Dialog is displayed. See Figure 5-41. This option is used to view a history log of drive error conditions. Note This command is not yet available. THIS FUNCTION IS NOT YET AVAILABLE Press Escape. Figure 5-41 Drive Error Log Dialog Drives Submenu When Drives is selected, the Drives Submenu is displayed. This submenu offers two selections. See Figure 5-42. The Drives Submenu defaults to the Drive State selection. >Drive State..
Drive State Dialog Note The library must be offline. When Drive State is selected, the Drive State Dialog is displayed. See Figure 5-43. This option indicates the current state of the selected drive. Drive: 1200< State: Loaded Clean Required: No [more] Figure 5-43 Drive State Dialog Step 1 Press the button to select the current parameter value and move to the Continuation Screen. — or — Press the button or the button to toggle between the Drive parameter value in the following list.
The value of the Cleaning Required parameter is the Drive State Dialog is supplied by the library firmware and can not be changed. The Cleaning Required parameter indicated if the drive requires cleaning (Yes, No). The Continuation Screen appears. See Figure 5-44.
Media Info Dialog When Media Info is selected, the Media Info Dialog is displayed. See Figure 5-45. This option shows media information in loaded drive elements. Enter SOURCE Coord: D 01< 1 A 01 OR Element: 01200 Accept: N Figure 5-45 Element Dialog Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list.
Press the button to confirm the changes. If Y is selected for the Accept parameter value, the Response Screen appears. See Figure 5-46. Otherwise, the changed parameters continue to display but no action is taken. Type: 6GB DLTIII Free: 00000000 KB Write Protected : No Mounts: 000016 Figure 5-46 01 Aug 2000 Response Screen Parameters Values Type Indicates the type of drive media (for DLT: NO TAPE, UNKNOWN, CLEAN TAPE, TK50 R0, TK70 R0, 2.
Commands Menu Selection The Commands Menu is primarily used to access various Scalar 1000 commands. Commands is selected from the Main Menu by following this procedure. Step 1 If necessary, press the Menu appears. button until the Main The Main Menu appears with the indicator positioned at Mode. Step 2 Press the button or the button to position the selection indicator at Commands. Step 3 Press the button. The Commands Menu appears. This menu offers four selections. See Figure 5-47.
Park When Park is selected, the gripper retracts and the accessor is immediately returned to the home position. The two Response Screens, see Figure 5-48 and Figure 5-49, appear. Requested command is now in progress ... Figure 5-48 Response Screen Requested command is now complete. Figure 5-49 Response Screen Move Submenu Note The Scalar 1000 must be Offline and Ready to use the commands under the Move Menu. When Move is selected, the Move Menu is displayed. This menu offers two selections.
Step 1 Press the Dialog. button to select the Position to Elem — or — Press the button or the Move Media Dialog. Press the button to select the button. The selected menu appears. Refer to Position to Elem Dialog on page 5-45 or Move Media Dialog on page 5-47.
Position to Elem Dialog When Position to Elem is selected, the Position to Elem Dialog is displayed. See Figure 5-51. This option moves the accessor in front of a specific element. Enter TARGET Coord: S< 01 2 A 01 OR Element : 00000 Accept: N Figure 5-51 Position to Elem Dialog Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list.
Accept Press the Step 2 Y to accept changes N to reject changes button to confirm the changes. Repeat Step 1 for all of the parameters in the list. If Y is selected for the Accept parameter value, two Response Screens appear. See Figure 5-52 and Figure 5-53. Otherwise, the changed parameters continue to display but no action is taken. Requested command is now in progress ... Figure 5-52 Response Screen The accessor positions itself in front of the specific element. Requested command is now complete.
Move Media Dialog When Move Media is selected, the Move Media Dialog is displayed. See Figure 5-54. This option moves cartridges between elements without host intervention. Enter SOURCE Coord: S< 02 1 A 01 OR Element: 00000 Accept : N Figure 5-54 Move Media Dialog Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list.
Accept Press the Step 2 Y to accept changes N to reject changes button to confirm the changes. Repeat Step 1 for all of the parameters in the list. If Y is selected for the Accept parameter value, the Target Menu appears. See Figure 5-55. Otherwise, the changed parameters continue to display but no action is taken. Enter TARGET Coord: I< 01 1 A 01 OR Element: 01182 Accept : N Figure 5-55 Target Dialog Step 3 Press the button to select the current parameter value and move to the next parameter.
The fifth field indicates the row of the column. (01 .. 12 for half inch and DLT coordinates) (01 .. 18 for 8mm coordinates) Element Indicates the element number which corresponds with the coordinate parameter Accept Y to accept changes N to reject changes Press the Step 4 button to confirm the changes. Repeat Step 3 for all of the parameters in the list. If Y is selected for the Accept parameter value, the Response Screens appear. See Figure 5-56 and Figure 5-57.
-=> WARNING <=The library may not be inventoried. Continue : N Figure 5-58 Warning Dialog Step 5 Press the button to select the current parameter value. — or — Press the button or the button to toggle between the parameter values. Parameters Values Continue Y to continue N to return to the previous menu If Y is selected for the Continue parameter, the Warning message is removed.
Inventory Dialog When Inventory is selected, the Inventory Dialog is displayed. See Figure 5-59. This option allows inventory of specified elements. Starting Inventory Coord: S 02 1< A 01 OR Element: 00000 Accept : N Figure 5-59 Inventory Dialog Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list.
Accept Y to accept changes N to reject changes Press the Step 2 button to confirm the changes. Repeat Step 1 for all of the parameters in the list. If Y is selected for the Accept parameter value, the Number of Elements Dialog, see Figure 5-60 appears. Otherwise, the changed parameters continue to display but no action is taken.
Requested command is now in progress ... Figure 5-61 Response Screen The barcode scanner inventories the specified storage cells. Requested command is now complete. Found : 100 Figure 5-62 01 Aug 2000 Response Screen Parameters Values Found Indicates the total number of cartridges detected.
Insert/Eject Submenu Note Use this option to insert or eject cartridge(s) without host intervention. When Insert/Eject is selected, the Insert/Eject Submenu is displayed. This submenu offers four selections. See Figure 5-63. The Insert/Eject Submenu defaults to the Insert selection. >Insert Insert Clean Tape... Eject... Eject Clean Tape... Figure 5-63 Insert/Eject Submenu To choose one of the four selections, follow the steps. Step 1 Press the button to select the Insert Screen.
Insert Screen When Insert is selected, the Insert Screen is displayed. See Figure 5-64. This option moves all cartridges found in the Insert/Eject station to the first available empty storage cells. If there are no cartridges in the Insert/Eject station, the response screen appears. See Figure 5-65. Insert in progress Source: 1182 Target: 0000 Figure 5-64 Insert Screen The accessor moves the cartridge(s) from the Insert/Eject station to the first available empty storage cell(s).
Insert Clean Tape Dialog When Insert Clean Tape is selected, the Insert Clean Tape Dialog is displayed. See Figure 5-66. This option moves selected cleaning cartridges from the Insert/Eject station to the designated empty storage cells. Enter SOURCE Coord: I 01< 1 A 01 OR Element: 01182 Accept: N Figure 5-66 Insert Clean Tape Dialog Step 1 Press the button to select the current parameter default value and move to the next parameter. — or — Press the button or the the Element parameter values.
Element Indicates the Insert/Eject station element number where the insert operation starts Accept Y to accept changes N to reject changes Press the button to confirm the changes. If Y is selected for the Accept parameter value, the Insert Range Dialog appears. See Figure 5-67. Otherwise, the changed actions continue to display but no action is taken.
Enter TARGET Coord: S 01 2< A 01 OR Element: 00000 Accept : N Figure 5-68 Target Dialog Step 3 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list.
If Y is selected for the Accept parameter value, the Usage Dialog appears. See Figure 5-69. If N is selected, the changed parameters continue to display but no action is taken. Current Usage: 000< Maximum Usage: 000 Accept: N Figure 5-69 Usage Dialog Step 4 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list.
Requested command is now complete.
Eject Dialog When Eject is selected, the Eject Dialog is displayed. See Figure 5-71. This option allows removal of cartridges without opening the front door(s) or without host intervention. The final destination is a slot in the Insert/Eject station. Enter SOURCE Coord: S< 02 1 A 01 OR Element : 01182 Accept: N Figure 5-71 Eject Dialog Step 1 Press the button to select the current parameter default value and move to the next parameter.
Element Indicates the element number which corresponds with the coordinate parameter Accept Y to accept changes N to reject changes Press the Step 2 button to confirm the changes. Repeat Step 1 for all of the parameter in the list. If Y is selected for the Accept parameter value, the Eject Screen appears. See Figure 5-72. Otherwise, the changed actions continue to display but no action is taken.
Eject Clean Tape Dialog When Eject Clean Tape is selected, the Eject Clean Tape Dialog is displayed. See Figure 5-74. This option allows removal of the cleaning cartridge(s) without opening the front door(s) or without host intervention Enter SOURCE Coord: S 01 1< A 13 OR Element: 00000 Accept: N Figure 5-74 Eject Clean Tape Dialog Step 1 Press the button to select the current parameter default value and move to the next parameter.
Element Indicates the element number which corresponds with the coordinate parameter Accept Y to accept changes N to reject changes Press the button to confirm the changes. If Y is selected for the Accept parameter value, the Clean Media Eject Screen appears. Refer to Figure 5-75. Otherwise, the changed actions continue to display but no action is taken.
Database Menu Selection The Database Menu is primarily used to gain element database information. Database is selected from the Main Menu by following this procedure. Step 1 If necessary, press the Menu appears. button until the Main The Main Menu appears with the indicator positioned at Mode. Step 2 Press the button or the button to position the selection indicator at Database. Step 3 Press the button. The Database Menu appears. This menu offers four selections. See Figure 5-77.
Media Dialog When Media is selected, the Media Dialog is displayed. This option obtains information about specific cartridges based on the barcode label. In default or mixed media mode, 6 characters are required. In extended mode, five characters are required. However, up to sixteen characters may be entered. Figure 5-78 shows the Media Dialog. Enter volume label: AAAAAA ^ Accept : N Figure 5-78 Media Dialog Step 1 Press the button to select the current parameter value and move to the next parameter.
BC : 000389 Type : 3590 Home : 0060 Current: 0060 Figure 5-79 Response Screen Parameters Values BC Indicates the value specified in the Media Dialog Type Indicates the cartridge media type (8590, DLT, 3480, 3490, 3590, 3590 DL, NCTP, 8mm) Home Indicates the home location of the cartridge Current Indicates the current location of the cartridge Element Dialog When Element is selected, the Element Dialog is displayed. See Figure 5-80.
Press the button or the button to toggle between the parameter values in the list. Parameters Values Coord The first field indicates the type of cell (I for I/E Station, D for Drive, S for Storage cell) The second field indicates the module number which can be modified if expansion modules are present (1 .. 4) The third field indicates the storage cell section (1 .. 4), drive bay (1 .. 2), or IE station number The fourth field indicates the column of the section (A .. E), drive port (A ..
Type : DLT IE Coord : I 01 1 A 12 Address : 0027 Index : 1193 [more] Figure 5-81 Response Dialog Step 3 Parameters Values Type Indicates the element type (3480 IE, DLT IE, MIXED IE, 8mm IE, 3480 STOR, DLT STOR, 8mm STOR, 3480 DRIVE, 3590 DRIVE, NCTP DRIVE, DLT4 DRIVE, DLT7 DRIVE, 8mm DRIVE) Coord Indicates the element coordinate location Address Indicates the current SCSI element address Index Indicates the database element index Press the Screen.
Config Dialog When Config is selected, the Config Dialog is displayed. See Figure 5-83. This option allows viewing of the current library configuration. Serial#: Frames : Cells : Drives : Figure 5-83 201100001 1 158 2 [more] Config Dialog Note If the Operator suspects a library configuration problem, use the Teach New option. Refer to Teach New Dialog on page 5-167. Step 1 Parameters Values Serial The serial number of the library Frames The number of frames (1 ..
ISSSS SSSSS SSSSS S3D2D Figure 5-84 Continuation Screen The screen is a graphics representation of the configuration of the library. Parameters Values Use the ISSSS button to scroll the cursor to the right. Press the Insert/Eject station (I) and 4 columns of Storage Cells (S) SSSSS 5 columns of Storage Cells (S) SSSSS 5 columns of Storage Cells (S) S3D2D 1 column of Storage Cells (S) and the number (1, 2, 3, 4, 6) of drives (D) in each port of the bottom drive bay.
Step 1 Press the button to select the Learn Dialog. — or — Press the button or the button to select the Manipulate Get Dialog, or Manipulate Put Dialog. Press the button. The selected menu appears. See Learn Dialog or refer to Manipulate Get on page 5-73 or Manipulate Put on page 5-74. Learn Dialog When Learn is selected, the Learn Dialog appears. See Figure 5-86. This option is reserved for Customer and Product Engineering Services and is password protected.
Manipulate Get When Manipulate Get is selected, the Manipulate Get Dialog appears. See Figure 5-87. This option is reserved for Customer and Product Engineering Services and is password protected. Enter password: [0000] ^ Accept: N Figure 5-87 Manipulate Get Dialog Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values. Parameters Values Password Selects the password (0000 ..
Manipulate Put When Manipulate Put is selected, the Manipulate Put Dialog appears. See Figure 5-88. This option is reserved for Customer and Product Engineering Services and is password protected. Enter password: [0000] ^ Accept: N Figure 5-88 Manipulate Put Dialog Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values. Parameters Values Password Selects the password (0000 ..
Setup Menu Selection The Setup Menu is primarily used to change operational characteristics of the Scalar 1000. Setup is selected from the Main Menu by following this procedure. Step 1 If necessary, press the Menu appears. button until the Main Step 2 Press the button or the button to position the selection indicator at Setup. Step 3 Press the button. The Setup Menu appears. This menu offers three selections. See Figure 5-89. The Setup Menu defaults to the Library selection. >Library Drives...
Library Submenu When Library is selected, the Library Submenu is displayed. This submenu offers seven selections. See Figure 5-90. The Library Submenu defaults to the SCSI selection. >SCSI Host... Trace... Security... Figure 5-90 Media... Emulate... RMU Advanced... Library Submenu To choose one of the seven selections, follow the steps. Step 1 Press the button to select the SCSI Submenu.
SCSI Submenu When SCSI is selected, the SCSI Submenu is displayed. This menu offers three selections. See Figure 5-91. The SCSI Submenu defaults to the Target ID selection. >Target ID... Parity... Ports... Figure 5-91 SCSI Submenu To choose one of the three selections, follow the steps. Step 1 Press the button to select the Target ID Dialog. — or — Press the button or the Parity Dialog or Ports Dialog. Press the button to select the button. The selected menu appears.
Target ID Dialog Note The Scalar 1000 defaults to SCSI ID 6. When Target ID is selected, the Target ID Dialog is displayed. See Figure 5-92. This option sets the SCSI ID of the library. Any change will not take effect until library power is cycled. Target ID: Bus 0: Bus 1: Figure 5-92 Next Now 4< 4 4 4 Accept: N Target ID Dialog Step 1 Press the button to select the current parameter value and move to the next parameter.
The value of the Bus 1 : Now parameter in the Target ID Dialog is supplied by the library firmware and can not be changed. The Bus 1 : Now parameter indicates the SCSI ID address that is currently in effect (0 .. 7) for bus port 1. If Y is selected for the Accept parameter value, the screen updates to reflect the changes. Otherwise, the changed parameters continue to display but no action is taken. Parity Dialog When Parity is selected, the Parity Dialog is displayed. See Figure 5-93.
and can not be changed. The Parity : Now parameter indicates if the SCSI parity is enabled (Yes, No). If parity is disabled, the retries parameter is ignored. The value of the Retries : Now parameter in the Parity Dialog is supplied by the library firmware and can not be changed. The Retries : Now parameter indicates the current number of retries (000 .. 255) allowed when a SCSI parity error is detected. Step 2 Repeat Step 1 for all of the parameters in the list.
Bus 1 : Type 01 Aug 2000 HVD indicates the bus is high voltage differential LVD indicates the bus is low voltage differential SE indicates the bus is single ended.
Host Dialog When Host is selected, the Host Dialog is displayed. See Figure 5-95. This option sets the type of host control. Host type: SCSI< Accept: N Figure 5-95 Host Dialog Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list. Note At present, only SCSI control is supported.
Trace Dialog When Trace is selected, the Trace Dialog is displayed. See Figure 5-96. This option defines and enables the serial service port. Trace output USE: NONE< Accept: N Figure 5-96 Trace Dialog Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter default values in the following list.
Security Dialog When Security is selected, the Security Dialog is displayed. See Figure 5-97. This option changes the Operator Panel LCD Security mode and/or the password that protects it. The Operator Panel can also be secured by the host. Whichever method is used to secure the Operator Panel must be the method used to release the security. If a password has not been set, the password dialog screen is shown.
If Y is selected for the Accept parameter and the security parameter is enabled, the Password Dialog appears. See Figure 5-98. If N is selected for the Accept parameter the changed parameters continue to display but no action is taken. Enter password : [0000] ^ Accept : N Figure 5-98 Password Dialog Step 3 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values.
Switching Security From On to Off To switch the security from on to off, follow the steps below: Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list. Parameters Values Security Y to enable LCD security N to disable LCD security Password If the password is to be changed, refer to Changing the Password on page 5-87.
Press the button or the button to toggle between the parameter values. Parameters Values Password Indicates the password (0000 .. 9999) Accept Y to accept changes N to reject changes Press the button to confirm the changes. If Y is selected for the Accept parameter value and the password is verified, the response screens appear. See Figure 5-101. Security is now OFF. Figure 5-101 Security is OFF Screen Step 4 Press the Submenu.
Password Y to change LCD password N to leave the LCD password as it is Accept Y to accept changes N to reject changes Press the Step 2 button to confirm the changes. Repeat Step 1 for all of the parameters in the list. If Y is selected for the Accept parameter and the password parameter, the Change Password Dialog appears. Otherwise, the changed parameters continue to display but no action is taken.
The password has been changed. Figure 5-103 Password Changed Step 4 01 Aug 2000 Press the Submenu.
Media Dialog When Media is selected, the Media Dialog is displayed. See Figure 5-104. This option controls the setting for media handling and reporting. The host software must support mixed media types if the library is operating in mixed media mode. For additional information, refer to document number 600840 Scalar 1000 SCSI Reference Manual.
Note Extend status, ASCQ, and VolTag are valid only when mixed media types are enabled and the cartridges are properly labeled with barcode labels having a media identifier.
Emulate Dialog When Emulate is selected, the Emulate Dialog is displayed. See Figure 5-105. This option changes the operating mode of the library. Different emulation modes cause the library to respond to SCSI INQUIRY commands in a manner consistent with the listed library emulation type. Any change will not take effect until the library power is cycled.
RMU Submenu The Scalar 1000 is equipped with a Remote Management Unit (RMU) which allows web-based library management via the Ethernet port. The RMU Submenu allows the administrator to set the initial values of the RMU network parameters to allow for remote access. When RMU is selected, the RMU Submenu is displayed. See Figure 5-106. >Name/DHCP... IP... Figure 5-106 RMU Submenu To choose one of the two selections, follow the steps. Step 1 Press the Dialog.
DHCP : N< Name : webfoot3 Accept : N Figure 5-107 Name/DHCP Dialog Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list. Parameters Values DHCP Y to enable DHCP communication N to disable the DHCP communication Name Indicates the host name assigned to the RMU for communicating with the library via the RMU (letters A .. Z and numbers 0 .. 9).
IP Dialog When IP is selected, the IP Dialog is displayed. See Figure 5-108. The IP Dialog allows the user to set the IP, Subnet and Gateway network addresses for library communication via the RMU. IP : 100<100.100.100 Sub: 225.225.225.225 Gat: 100.100.100.100 Accept: N Figure 5-108 IP Dialog Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list.
Advanced Dialog When Advanced is selected, the Advanced Dialog is displayed. See Figure 5-109. This option enables or disables the automatic configuration, calibration, and cartridge scan when the library has power cycled. If Auto Inventory is disabled, the element status is not known until the host issues an Initialize Status command or an Inventory is performed via the Operator Panel. If Auto Inventory is enabled, an automatic cartridge inventory is executed on each power cycle.
Operating Mode 0 to select Normal mode of operation 1 to select a 3 second delay before a Get operation is performed on a DLT drive after detecting that a tape was ejected 2 to allow the Scalar 1000 to issue an UNLOAD command to the Model 4001S, Model 7001S, Model 8001S, AIT3102, AIT5002, AIT5102, and NCTP drives if the cartridge is not ejected by the host 3 to activate options 1 and 2 4 to disable automatic cartridge recovery on PUT failures 5 to activate options 1 and 4 6 to activate options 2 and 4 7 to
Drives Dialog When Drives is selected, the Drives Dialog is displayed. See Figure 5-110. This options allows modification of the drive SCSI ID for the Model 4001S, Model 7001S, Model 8001S, AIT3102, AIT5002, AIT5102, and NCTP drives. Drive ID Override: Enabled < Set Drive ID: N Accept: Figure 5-110 N Drives Dialog Step 1 Press the button to select the current parameter value and move to the next parameter.
If Drive ID Override is disabled and Y is selected for Accept, the changed parameters continue to display but no new screen action is displayed. — or — If N is selected for the Accept parameter, the changed parameters continue to display but no action is taken. Drive 1200< Figure 5-111 SCSI ID Next Now 01 01 Accept: N SCSI ID Dialog Step 2 Press the button to select the current parameter value and move to the next parameter.
Cleaning Submenu When Cleaning is selected, the Cleaning Submenu is displayed. This submenu offers two selections. See Figure 5-90. The Cleaning Submenu defaults to the Drives selection. >Drives... Media... View... Figure 5-112 Cleaning Submenu To choose one of the three selections, follow the steps. Step 1 Press the button to select the Drives Dialog. — or — Press the button or the button to select the Media Dialog or View Dialog. Press the button. The selected menu appears.
Drives Dialog When Drives is selected, the Drives Dialog is displayed. See Figure 5-113. This option allows selection of automatic cleaning for Model 4001S, Model 7001S, Model 8001S, AIT3102, AIT5002, AIT5102, and NCTP drives. AutoClean: Schedule: Hours: 00 Figure 5-113 Y< Delayed Min :00 Accept:: N Drives Dialog Step 1 Press the button to select the current parameter value and move to the next parameter.
If Y is selected for the Accept parameter value, the screen updates to reflect the changes. Otherwise, the changed parameters continue to display but no action is taken.
Media Dialog When Media is selected, the Media Dialog is displayed. See Figure 5-114. This option allows an operator to specify DLT or AIT 8mm cleaning tapes by BC in the library. A BC mask is a valid, partial barcode label that can be followed by a wildcard character (*). Barcode labels that match the mask are moved as cleaning tapes. The BC mask must not match more than a maximum of 127 cleaning tapes.
Current Use: Maximum Use: 000< 000 Accept: N Figure 5-115 Usage Dialog Step 3 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list. Note If the BC mask selects more than 1 cleaning tape, the current and maximum usage parameter values are applied to each cleaning tape.
View Dialog When View is selected, the View Dialog is displayed. See Figure 5-116. This option allows an operator to see a specific cleaning tape usage. BC: CLN001 Status: Reset Max: 000 Count: 000 Index: 001
Press the Step 2 button to confirm the changes. Repeat Step 1 for all of the parameters in the list. If Y is selected for the Reset parameter value, the changed parameter values are accepted. Otherwise, the changed parameters continue to display but no action is taken.
Utils Menu Selection The Utils Menu is primarily used to execute various Scalar 1000 utilities. Utils is selected from the Main Menu by following this procedure. Step 1 If necessary, press the Menu appears. button until the Main The main Menu appears with the indicator positioned at Mode. Step 2 Press the button or the the selection indicator at Utils. Step 3 Press the button to position button. The Utils Menu appears. This menu offers two selections. See Figure 5-117.
Library Submenu When Library is selected, the Library Submenu is displayed. This submenu offers five selections. See Figure 5-118. The Library Submenu defaults to the Screen selection. >Screen... Audio... Dump Time... Figure 5-118 Date... Library Menu To choose one of the five selections, follow the steps. Step 1 Press the button to select the Screen Dialog. — or — Press the button or the button to select Audio Dialog, Dump, Time Dialog, or Date Dialog. Press the button.
Screen Dialog When Screen is selected, the Screen Dialog is displayed. See Figure 5-119. This option controls the settings related to the physical Operator Panel LCD screen. Backlight Enable: Use Screen Saver: Y< N Accept: N Figure 5-119 Screen Dialog Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list.
Timeout Value Password Enable : 30< : N Accept: N Figure 5-120 Timeout Value Dialog Step 3 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list.
Step 5 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values. Parameters Values Password Selects the password (0000 .. 9999) Press the button to confirm the changes. Step 6 Repeat Step 5 for all password parameters. Step 7 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values.
incorrectly, the following Invalid Dialog appears. See Figure 5-123. You did not enter a valid password Figure 5-123 Invalid Password Dialog Step 8 5-112 Menus and Commands Press the Dialog.
Audio Dialog When Audio is selected, the Audio Dialog is displayed. See Figure 5-124. This option controls the settings to enable or disable the audio alarm. Audio Enabled : Y< Accept : N Figure 5-124 Audio Dialog Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list.
Dump When Dump is selected, the Response Screen is displayed after the dump execution is complete. See Figure 5-125. This option captures the current operating state of the library which consists of database, log, parameter, and trace information. The dump file may be retrieved by a SCSI READ BUFFER command. The code dump is now completed.
Time Dialog When Time is selected, the Time Dialog is displayed. See Figure 5-126. This option sets the 24 hour time format of the library. The library does not automatically adjust for Daylight Savings Time. Hour Min Sec : 18< : 24 : 25 Accept: N Figure 5-126 Time Dialog Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list.
Date Dialog When Date is selected, the Date Dialog is displayed. See Figure 5-127. This option sets the date displayed by the library. Year : 00< Month : 08 Day : 16 Accept: N Figure 5-127 Date Dialog Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list. Parameters Values Year Indicates the last two digits of the year (00 ..
Drives Submenu When Drives is selected, the Drives Submenu is displayed. This submenu offers four selections. See Figure 5-128. The Drives Submenu defaults to the Update Microcode selection. The menu selection supports only Model 4001S, Model 7001S, Model 8001S, AIT3102, AIT5002, AIT5102, and NCTP drives. >Update Microcode... Clean Drives... Unload Drives... Initialize Figure 5-128 Drives Submenu To choose one of the four selections, follow the steps.
Update Microcode Dialog When Update Microcode is selected, the Update Microcode Dialog is displayed. See Figure 5-129. This option initiates a drive firmware update. The operator must determine the location coordinate or element address of the firmware update tape prior to following this procedure. Enter SOURCE Coord: I 01 1< A 12 OR Element: 01193 Accept : N Figure 5-129 Update Microcode Dialog Step 1 Press the button to select the current parameter default value and move to the next parameter.
column The fifth field indicates the row of the column. (01 .. 12 for half inch and DLT coordinates) (01 .. 18 for 8mm coordinates) Element Indicates the element number which corresponds with the coordinate parameter Accept Y to accept changes N to reject changes Press the Step 2 button to confirm the changes. Repeat Step 1 for all of the parameter in the list. If Y is selected for the Accept parameter value, the Element Dialog appears. See Figure 5-130.
drive bay (1 .. 2), or IE station number The fourth field indicates the column of the section (A .. E), drive port (A .. B) or IE station column The fifth field indicates the row of the column. (01 .. 12 for all half inch and DLT coordinates) (01 .. 18 for 8mm coordinates) Element Indicates the element number which corresponds with the coordinate parameter Accept Y to accept changes N to reject changes Press the Step 4 button to confirm the changes. Repeat Step 3 for all of the parameter in the list.
Parameters Values Drive Range Indicates the number of drives to receive the firmware update. Drives are in order of their element address Accept Y to accept changes N to reject changes Press the Step 6 button to confirm the changes. Repeat Step 5 for all of the parameter in the list. If Y is selected for the Accept parameter value, the two Response Screen appears. See Figure 5-132. Otherwise, the changed actions continue to display but no action is taken.
Cancel Press the Y to cancel the cycle N to continue the cycle button to confirm the changes. If Y is selected for the Cancel parameter, the cycle is canceled and the Cancel parameter does not appear.
Clean Drives Dialog When Clean Drives is selected, the Clean Drives Dialog is displayed. See Figure 5-133. This option initiates a drive cleaning operation. Enter TARGET Coord: D< 01 1 A 01 OR Element: 01200 Accept: N Figure 5-133 Clean Drives Dialog Step 1 Press the button to select the current parameter default value and move to the next parameter. — or — Press the button or the button to toggle the Coordinate parameter values.
Accept Press the Step 2 Y to accept changes N to reject changes button to confirm the changes. Repeat Step 1 for all of the parameter in the list. If Y is selected for the Accept parameter value, the Progress Screen appears, see Figure 5-134, followed by the Response Screen, see Figure 5-135. Otherwise, the changed actions continue to display but no action is taken. Requested command is now in progress ... Figure 5-134 Progress Screen Requested command is now complete.
Unload Drives Dialog When Unload Drives is selected, the Unload Drives Dialog is displayed. See Figure 5-136. This option initiates a tape eject operation from the specified drive. Enter SOURCE Coord: D 01< 1 A 01 OR Element: 01200 Accept: N Figure 5-136 Unload Drives Dialog Step 1 Press the button to select the current parameter default value and move to the next parameter. — or — Press the button or the parameter values.
Accept Press the Step 2 Y to accept changes N to reject changes button to confirm the changes. Repeat Step 1 for all of the parameter in the list. If Y is selected for the Accept parameter value, the Progress Screen appears, see Figure 5-137, followed by the Response Screen, see Figure 5-138. Otherwise, the changed actions continue to display but no action is taken. Requested command is now in progress ... Figure 5-137 Progress Screen Requested command is now complete.
Initialize Submenu When Initialize is selected, the Initialize Submenu is displayed. This submenu offers two selections. See Figure 5-139. The Initialize Submenu defaults to the Communication selection. >Communication SCSI Figure 5-139 Initialize Submenu To choose one of the two selections, follow the steps. Step 3 Press the button to select Communication. — or — Press the SCSI. button or the Press the button. button to select The selected screen appears.
Communication When Communication is selected, the Progress Screen, see Figure 5-140, followed by the Response Screen, see Figure 5-141, appear. This option initiates communication tests to drives and initializes the drive state of Model 4001S, Model 7001S, Model 8001S, AIT3102, AIT5002, AIT5102, and NCTP appropriately. Requested command is now in progress ... Figure 5-140 Progress Screen Requested command is now complete.
SCSI When SCSI is selected, all SCSI IDs are initialized to the SCSI IDs previously set by the operator. The Progress Screen, see Figure 5-141, followed by the Response Screen, see Figure 5-142, appear. Requested command is now in progress ... Figure 5-142 Progress Screen Requested command is now complete.
Service Menu Selection The Service Menu is primarily used to execute various Scalar 1000 diagnostics and exercisers. These options are primarily intended for trained service representatives. Service is selected from the Main Menu by following this procedure. Step 1 If necessary, press the Menu appears. button until the Main Step 2 Press the button or the button to position the selection indicator at Service. Step 3 Press the button. The Service Menu appears. This menu offers six selections.
Start Dialog When Start is selected, the Preventative Maintenance (PM) Due Dialog, see Figure 5-145, is displayed. If a reset of the PM values is selected, the Confirm Change Dialog, see Figure 5-146, is displayed. The menu flow continues with either Start Dialog with No Errors or Start Dialog with Errors. PM Due: 01/01/98 X Remain: 00500000 Y Remain: 000500000 Reset PM values? N< Figure 5-145 Preventive Maintenance Due Dialog Step 1 Press the value.
Step 2 Press the button to select the current parameter value. — or — Press the button or the button to toggle between the parameter values in the following list. Parameters Values Confirm change Y to accept changes to PM values. N to reject changes to PM values. Press the button to confirm the changes. If Y is selected for the Confirm Change parameter value and no errors were present, the Start Dialog with No Error, see Figure 5-147, is displayed.
Found SAC at: 03/05/97 14:04:38.68 SAC : E1 Error: 0x02030204 Figure 5-148 Start Dialog with Error Note Press the button or the button to scroll between SAC values. Diags Submenu Note Diagnostics are grouped according to functional areas. When Diags is selected, the Diags Submenu is displayed. This menu offers six selections. See Figure 5-149. The Diags Submenu defaults to the DI/DO selection. >DI/DO Gripper Scanner Accessor Figure 5-149 SelfTest... PowerSup...
Press the button or the button to select the Gripper Submenu, Scanner Submenu, Accessor Submenu, SelfTest Dialog, or PowerSup Dialog. Press the button. The selected menu appears. Refer to DI/DO Submenu on page 5-135, Gripper Submenu on page 5-143, Scanner Submenu on page 5-154, Accessor Submenu on page 5-159, SelfTest Dialog on page 5-164, or PowerSup Dialog on page 5-165.
DI/DO Submenu Note When DI/DO is selected, the DI/DO Submenu is displayed. This menu offers three selections. See Figure 5-150. The DI/DO Submenu defaults to the Loopback selection. Use these tests to isolate problems in the system. >Loopback... Sensors Locks... Figure 5-150 DI/DO Submenu To choose one of the three selections, follow the steps. Step 1 Press the button to select the Loopback Dialog. — or — Press the button or the button to select the Sensors Dialog or Locks Dialog. Press the button.
Loopback Dialog When Loopback is selected, the Loopback Dialog is displayed. See Figure 5-151. This option sends test signals from the main control card to other system cards and back. Device: LMC Number: 01 Results: Accept: N Figure 5-151 Loopback Dialog Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list.
Device: LMC Number: 01 Results: PASSED Accept: N Figure 5-152 01 Aug 2000 Response Dialog Parameters Values Results Passed indicates that the test executed successfully Failed indicates that the test did not execute successfully Accept Y to accept change N to reject changes Service Menu Selection 5-137
Sensors Submenu When Sensors is selected, the Sensors Submenu is displayed. This menu offers two selections. See Figure 5-154. The Sensors Submenu defaults to the Wrap selection. >Wrap ... Real Time ... Figure 5-153 Sensors Submenu To choose one of the two selections, follow the steps. Step 1 Press the button to select the Wrap Dialog. — or — Press the button or the Real Time Dialog. Press the button to select the button. The selected menu appears.
Wrap Dialog When Wrap is selected, the Wrap Dialog is displayed. See Figure 5-154. This option tests the ability of the system sensors to report state changes. Device: ALL SENSORS Results : < Accept: N Figure 5-154 Wrap Dialog Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list.
Device: ALL < Results : Passed Accept: N Figure 5-155 5-140 Menus and Commands Response Dialog Parameters Values Results Passed indicates that the test executed successfully Failed indicates that the test did not execute successfully Accept Y to accept changes N to reject changes 600561-A
Real Time When Real Time is selected, the Real Time screen is displayed. See Figure 5-156. This option indicates the state changes of the sensors and allows real time viewing of the state changes.
Locks Dialog When Locks is selected, the Locks Dialog is displayed. See Figure 5-157. This option exercises the I/E station lock without having to open and close the station. If media removal has been prevented by the host, the I/E station can not be unlocked. Pressing Enter will toggle the lock. State : Unlocked Figure 5-157 Locks Dialog Step 1 5-142 Menus and Commands Press the button to select the State parameter.
Gripper Submenu When Gripper is selected, the Gripper Submenu is displayed. This menu offers two selections. See Figure 5-158. The Gripper Menu defaults to the Get/Put Storage selection. >Get/Put Storage... Get/Put Drives... Step... Figure 5-158 Gripper Submenu To choose one of the three selections, follow the steps. Step 1 Press the Dialog. button to select the Get/Put Storage — or — Press the button or the button to select the Get/Put Drives Dialog or Step Dialog Press the button.
Get/Put Storage Dialog When Get/Put Storage is selected, the Get/Put Storage Dialog is displayed. See Figure 5-159. This option does an in place GET and PUT of all cartridges in the specified range. No element to element movements are allowed. The gripper test will get/put media Cycles to run: 009< [more] Figure 5-159 Get/Put Storage Dialog Step 1 Press the button to select the current parameter value and move to the next parameter.
Step 3 Press the button to select the current parameter default value and move to the next parameter. — or — Press the button or the button to toggle the Coordinate parameter values. Parameters Values Coord The first field indicates the type of cell (I for I/E Station, D for Drive, S for Storage cell) The second field indicates the module number which can be modified if expansion modules are present (1 .. 4) The third field indicates the storage cell section (1 .. 4), drive bay (1 ..
Number of Elements to Get/Put : 0100 Accept : N Figure 5-161 Number of Elements Dialog Step 5 Press the button to select the current parameter default value and move to the next parameter. — or — Press the button or the the parameter values. Parameters Values Get/Put Selects the number of elements where get/put operations are applied (0000 .. 1181) Accept Y to accept changes N to reject changes Press the Step 6 button to toggle button to confirm the changes.
Step 7 Note If the operation stops, the Cancel parameter no longer displays. Press the button or the Cancel parameter values. button to toggle the Parameters Values Cycle The number of completed test cycles of the requested cycles Status Shows the status of the indicated cycle (Running, Completed, Error, Canceled) Cancel Y to cancel the test. N to continue running the test. Press the button to confirm the changes. If Y is selected for the Cancel parameter value, the test is cancelled.
Get/Put Drives Dialog When Get/Put Drives is selected, the Get/Put Drives Dialog is displayed. See Figure 5-163. This option does an in place GET and PUT of all cartridges in the specified range to the specified drive(s). The menu selection supports only Model 4001S, Model 7001S, Model 8001S, AIT3102, AIT5002, AIT5102, and NCTP drives.
Enter SOURCE Coordinate : D 01 1< A 01 OR Element : 01200 Accept : N Figure 5-164 Continuation Dialog Step 3 Press the button to select the current parameter default value and move to the next parameter. — or — Press the button or the button to toggle the Coordinate parameter values.
Press the Step 4 button to confirm the changes. Repeat Step 3 for all of the parameter in the list. If Y is selected for the Accept parameter value, the Number of Elements Dialog appears. See Figure 5-165. Otherwise, the changed parameters continue to display but no action is taken. Number of Drives to Get/Put : 0010< Accept : N Figure 5-165 Number of Elements Dialog Step 5 Press the button to select the current parameter default value and move to the next parameter.
DIAGS GRIPPER TEST Cycle : 006 of 009 Status: Running Cancel: N< Figure 5-166 Response Dialog Step 7 Note If the operation stops, the Cancel parameter no longer displays. Press the button or the Cancel parameter values. button to toggle the Parameters Values Cycle The number of completed test cycles of the requested cycles Status Shows the status of the indicated cycle (Running, Completed, Error, Canceled) Cancel Y to cancel the test. N to continue running the test.
Step Dialog When Step is selected, the Step Dialog is displayed. See Figure 5-167. This option provides low level control of Gripper functions. The function should be used by trained service personnel. This test should be used by trained personnel ONLY! Continue: N< Figure 5-167 Step Dialog Step 1 Press the button to select the current parameter value. — or — Press the button or the button to toggle between the parameter values. Parameters Values Continue Y to continue the test.
Press the button or the button to toggle between the parameter values in the following list. Parameters Values Grip Open indicates that the Gripper is open Closed indicates that the Gripper is closed Reach Retracted indicates that the Gripper is retracted Extended indicates that the Gripper is extended Speed Normal uses predefined normal speed Slow uses predefined slow speed Press the Step 3 01 Aug 2000 button to confirm the changes. Repeat Step 1 for the desired parameters in the list.
Scanner Submenu When Scanner is selected, the Scanner Submenu is displayed. This submenu offers two selections. See Figure 5-169. The Scanner Menu defaults to the Adjust selection. >Adjust... Triger... Fiducial Test... Figure 5-169 Scanner Submenu To choose one of the two selections, follow the steps. Step 1 Press the button to select the Adjust Dialog. — or — Press the button or the Trigger Dialog. Press the button to select the button. The selected menu appears.
Adjust Dialog When Adjust is selected, the Adjust Dialog is displayed. See Figure 5-170. This option positions the barcode scanner beam over a specific location. This option is used by Customer Engineer to adjust the beam. Refer to the Scalar 1000 Maintenance Guide for the step by step procedure to adjust the scanner. DIAGS SCANNER ADJUST Adjust scanner until beam is level. Y Position : 1002.0< Figure 5-170 Adjust Dialog Step 1 01 Aug 2000 Press the button or the Y position parameter values.
Trigger Dialog When Trigger is selected, the Trigger Dialog is displayed. See Figure 5-171. This option triggers the barcode scanner to read, decode, and display whatever barcode label is within the range of the beam. SCANNER TRIGGER Pressing Enter will trigger scanner BC :000389 Figure 5-171 Response Screen Step 1 5-156 Menus and Commands Press the button to trigger the scanner. Parameters Values BC Returns the barcode label information for the storage cartridge.
Fiducial Test Dialog When Fiducial Test is selected, the Fiducial Test Dialog is displayed. See Figure 5-172. This option allows the operator to test the scanner to determine if it is functioning properly. Loops: 00 Test: 1 Xcnt: 000000 Ycnt: 000000 Continue: Y< Figure 5-172 Fiducial Test Dialog Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list.
not be changed. The YCnt parameter indicates the Y edge position count in tenths of a mm. Step 2 5-158 Menus and Commands Repeat Step 1 several times to ensure the X and Y position counts are within 10 tenths of a mm each time the test is run.
Accessor Submenu When Accessor is selected, the Accessor Submenu is displayed. This menu offers two selections. See Figure 5-173. The Accessor Submenu defaults to the Move selection. >Move... Step... Figure 5-173 Accessor Submenu To choose one of the two selections, follow the steps. Step 1 Press the button to select the Move Dialog. — or — Press the button or the Step Dialog. Press the button to select the button. The selected menu appears.
Move Dialog When Move is selected, the Move Dialog is displayed. See Figure 5-174. This options moves the accessor in a figure eight pattern without accessing cartridges. The accessor test does corner moves. Cycles to run: 009< Accept: N Figure 5-174 Move Dialog Step 1 Press the button to select the current parameter value and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list.
DIAGS ACCESSOR TEST Cycle : 006 of 009 Status: Running Cancel: N< Figure 5-175 Response Dialog Step 3 Note If the operation stops, the Cancel parameter no longer displays. Press the button or the Cancel parameter values. button to toggle the Parameters Values Cancel Y to cancel the test. N to continue running the test. Press the button to confirm the changes. The value of the Cycle parameter in the Response Dialog is supplied by the library firmware and can not be changed.
Step Dialog When Step is selected, the Step Dialog is displayed. See Figure 5-176. This option provides low level control of the accessor. The option should be used by trained service personnel. This test should be used by trained personnel ONLY! Continue: N< Figure 5-176 Step Dialog Step 1 Press the button to select the current parameter value. — or — Press the button or the button to toggle between the parameter values. Parameters Values Continue Y to continue the test.
Parameters Values Press the X axis Indicates X axis location button or the Y axis Indicates Y axis location Amount Indicates move step distance (1, 10 .. 100 mm) Note button to move the selected axis by the specified amount. Use the button to move between the parameters.
SelfTest Dialog When SelfTest is selected, the SelfTest Dialog is displayed. See Figure 5-178. This option executes a predetermined sequence of diagnostics and exercisers. The self test runs a sequence of diags. Cycles to run: 009< Accept: N Figure 5-178 SelfTest Dialog Step 1 Press the button to select the current parameter values and move to the next parameter. — or — Press the button or the button to toggle between the parameter values in the following list.
Step 3 Note If the operation stops, the Cancel parameter no longer displays. Press the button or the Cancel parameter values. button to toggle the Parameters Values Cycle The number of completed test cycles of the requested cycles Status Shows the status of the indicated cycle (Running, Completed, Error, Canceled) Cancel Y to cancel the test. N to continue running the test. Press the button to confirm the changes. If Y is selected for the Cancel parameter value, the test is cancelled.
Step 1 5-166 Menus and Commands Press the Submenu.
Teach Submenu When Teach is selected, the Teach Submenu is displayed. This submenu offers two selections. See Figure 5-182. The Teach menu defaults to the Teach New selection. >Teach New... Teach Current... Figure 5-182 Teach Submenu To choose one of the two selections, follow the steps. Step 1 Press the Dialog. button to select the Teach New — or — Press the button or the Teach Current Dialog. Press the button to select the button. The selected menu appears.
A teach new will clear and reset the complete database. Continue: N Figure 5-183 Teach New Dialog Step 1 Press the button to select the current parameter value. — or — Press the button or the button to toggle between the parameter values. Parameters Values Continue Y to continue the teach new. N to return to the previous menu If Y is selected for the Continue parameter value, the Progress Screen, see Figure 5-184, followed by the Response Screen, see Figure 5-185, appear.
The requested teach completed OK. Figure 5-185 Response Screen Teach Current Dialog When Teach Current is selected, the Teach Current Dialog is displayed. See Figure 5-186. This option re-calibrates the library coordinate information. All previous configuration and inventory information is retained. A teach current will reset the positional information only. Continue: N Figure 5-186 Teach Current Dialog Step 1 Press the value.
Starting Box: 01 Ending Box : 02 Accept: N Figure 5-187 Starting and Ending Dialog Step 2 Press the button to select the current parameter value. — or — Press the button or the button to toggle between the Starting Box and Ending Box parameters. Parameters Values Starting Box Indicates which frame to begin the Teach Current operation. (1 .. 4 depending on the number of expansion modules) Ending Box Indicates which frame to end the Teach Current operation. (1 ..
The requested teach is now in progress... Figure 5-188 Progress Screen The requested teach completed OK.
SAC Dialog When SAC is selected, the SAC Dialog is displayed. See Figure 5-190. This option displays explanations of the displayed Service Action Code. Enter SAC to lookup: SAC: E1 I/E station door is not closed Figure 5-190 SAC Dialog For additional information on the SAC codes, refer to Service Action Codes on page 7-3. Note Press the button or the button to scroll between SAC values. Demo Dialog When Demo is selected, the Demo Dialog is displayed. See Figure 5-191.
— or — Press the button or the button to toggle between the parameter values in the following list. Parameters Values Cycles to Run Indicates the desired cycle count (001 .. 999) Accept Y to accept changes N to reject changes Press the Step 2 button to confirm the changes. Repeat Step 1 for all of the parameters in the list. If Y is selected for the Accept parameter, the Include Drives Dialog appears. See Figure 5-192. Otherwise, the changed parameters continue to display but no action is taken.
If Y is selected for the Accept parameter value, the Response Dialog appears. See Figure 5-193. Otherwise, the changed parameters continue to display but no action is taken. Demo with Drives Cycle : 5 of 9 Status: Running Cancel: N< Figure 5-193 Response Dialog Step 5 Note If the operation stops, the Cancel parameter no longer displays. Press the button or the the Cancel parameter values. button to toggle Parameters Values Cancel Y to cancel the test. N to continue running the test.
Advanced Dialog When Advanced is selected, the Advanced Dialog is displayed. See Figure 5-194. >Status Change... Drive Mounts... Other Functions... Figure 5-194 Advanced Dialog To choose one of the two selections, follow the steps. Step 1 Press the Dialog. button to select the Status Change — or — Press the button or the button to select the Drive Mounts Dialog, or the Other Functions Dialog. Press the button. The selected menu appears.
Enter password: [0000] ^ Accept: N Figure 5-195 Status Change Dialog Drive Mounts Dialog When Drive Mounts is selected, the Drive Mounts Dialog is displayed. See Figure 5-196. This dialog is reserved for Customer and Product Engineering Services and is password protected. Enter password: [0000] ^ Accept: N Figure 5-196 Drive Mounts Dialog Other Functions Dialog When Other Functions is selected, the Other Functions Dialog is displayed. See Figure 5-197.
About Screen Selection The About screen displays information about the version of embedded software. About is selected from the Main Menu by following the procedure. Step 1 If necessary, press the Menu appears. button until the Main The Main Menu appears with the indicator positioned at Mode. Step 2 Press the button or the button to position the selection indicator at About. Step 3 Press the button. The About Screen appears. See Figure 5-198. FIRMWARE LEVEL (c) Copyright 2000 Version 3.00.
5-178 Menus and Commands 600561-A
6 Processing Media Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3 Inserting Media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3 Insert/Eject Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3 Direct Insertion into the Control or Expansion Module . . . . . . . . . . . . . . . . . .
6-2 Processing Media 600561-A
6- Overview During the course of normal operations, it may become necessary to add or remove storage cartridges to or from the library. Two methods are available to insert or eject storage cartridges. Information is also supplied about processing the media and using cleaning media. Inserting Media Placing new media into the library is accomplished by using the Insert/Eject Station. Alternately, the cartridge could be placed directly into a specific storage cell location by the Operator.
Direct Insertion into the Control or Expansion Module STATIC SENSITIVE There are two levels of access control. The first is a mechanical lock which is manipulated by the key. The second is password protection which is invoked by either the host SCSI Mode Select command, LCD Page Security option, or through the Security option on the Operator panel. To directly insert media, follow this procedure. If enabled, all locks must be disabled before the procedure will succeed.
Insert/Eject Station Using the Insert/Eject Station to remove cartridges does not disrupt library operations. To remove cartridges through the Insert/Eject Station, follow this procedure. Step 1 Issue a SCSI MOVE MEDIUM command from the host with the source storage cell and the destination Insert/Eject Station cell. — or — Execute the Eject command from the Operator panel. Refer to Eject Dialog on page 5-61. The Insert/Eject Station is locked while the accessor moves the cartridge.
Direct Removal from the Control or Expansion Module STATIC SENSITIVE There are two levels of access control. The first is a mechanical lock which is manipulated by the key. The second is password protection which is invoked by either the host SCSI Mode Select command, LCD Page Security option, or through the Security option on the Operator panel. To directly remove media, follow this procedure. If enabled, all locks must be disabled before the procedure will succeed.
Note The 6 character 8mm bar code labels contain a 7th character checksum, which is ignored in DEFAULT media mode, is reported as A in MIXED media mode, and interpreted as part of the bar code in EXTENDED mode. The labels may have 5-16 characters (library in EXTENDED media mode), or 6 characters plus a possible media identifier character (library in DEFAULT media mode), or 6 characters with an additional media identifier character (library in MIXED media mode).
• ANSI MH10.8M-1983 Standard • Number of digits: ♦ 5 - 16 in extended mode 6 (7 including media character) in default and mixed media mode Background reflection: at least 25 percent Print contrast: at least 75 percent Ratio: at least 2.2 Module: 250 mm Print tolerance: ± 57 mm ♦ • • • • • • Additional Requirements • • • • • • • Length of the rest zones: 5.25 mm ± 0.
Table 6-1 Supported Labels Media Label Source DLT-III Engineered Data Products 1703-0C DLT-IV Engineered Data Products 1703-0D DLT-IIIXT Engineered Data Products 1703-0E AIT Series (8mm) Engineered Data Products Barcode Label Restrictions Note Duplicate BC labels are not supported even if different media labels are used. The Scalar 1000 supports a range of media labels. With mixed media enabled, some additional restrictions apply.
1 2 Figure 6-1 3 4 5 6 7 8 9 Barcode Label Examples The numeric listing above each label type identifies the restriction associated with the label. 6-10 Processing Media 1 This is a 6 character Code 39 forward printed label for a half inch cartridge. This label can be used only when mixed media is disabled. 2 This is a 6 character Code 39 backward printed label for a half inch cartridge. This label can be used only when mixed media is disabled.
5 This is a 6 character Tri-Optic forward printed label with one additional character media identifier for a half inch cartridge. This label can be used in both mixed media enabled and mixed media disabled applications. 6 This is a 6 character Code 39 forward printed label for a DLT cartridge. This label can be used only when mixed media is disabled. 7 This is a 6 character Code 39 forward printed label for a DLT cartridge. This label can be used only when mixed media is disabled.
Note • It is possible to use the Insert/Eject station as storage for cleaning media but this practice is not recommended. Under the Commands Menu selection, Insert/Eject Submenu, and Clean Media Insert Dialog, move either a single media or bulk load cleaning media from the Insert/Eject station to the designated empty storage cells. Refer to Insert Clean Tape Dialog on page 5-56. As an alternate method, open the Scalar 1000 door(s) and manually bulk load the cleaning media into empty storage cells.
Since the Scalar 1000 will go NOTREADY for the duration of the cleaning operation, the Delayed option should be initiated during routine or normal downtime. • If the Immediate option is selected, every dismount causes the firmware to check for a cleaning request from the drive and, if requested, cleans the drive immediately. If the firmware receives a clean drive request, the dismount response is held until the cleaning process has finished.
6-14 Processing Media 600561-A
7 Error Messages Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3 Service Action Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3 Operator Intervention Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10 Error Log Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7-2 Error Messages 600561-A
7- Overview When a failure occurs, the Scalar 1000 firmware performs error recovery and reporting. If the failure requires a service call, Service Action Codes (SACs) are generated and posted to the Operator panel. When an operator intervention is required, a message is generate at the Operator Panel. Service Action Codes Table 7-1 lists the SACs and the corresponding actions that can be performed by an Operator.
Table 7-1 Reported SAC 03 SAC Reporting Perform these actions Type 3 Software errors including Microcode and Operating System errors. Retrieve the current Microcode dump. Force another Microcode Dump and retrieve it. Power Off and On the Scalar 1000 Control Module to recover from the error. Use the Circuit Breaker to perform this action. Wait at least one minute between Power Off and Power On. If the problem is not corrected or persists, contact ATAC.
Table 7-1 Reported SAC 13 SAC Reporting Perform these actions Cannot read Cartridge Barcode label. Check Cartridge label to ensure that it meets specifications and is installed properly. Look for the following: • a damaged or dirty label • Library not configured for correct media mode • too short (less than 5 characters) or too long (more than 16 characters) label. The cell in question is displayed with the Service Action Code. Run the Start Option from the Operator Panel to retrieve this information.
Table 7-1 SAC Reporting Reported SAC 33 Perform these actions Wrong SCSI bus connection is detected. Verify that the Host SCSI bus is the same type as the library SCSI Adapter card by using the Operator Panel Main Menu. If the problem is not corrected or persists, contact ATAC. 40 The Library Aisle power cannot be enabled. Check that all doors are closed. Contact ATAC. 50 A cartridge is not properly seated in the storage cell.
Table 7-1 SAC Reporting Reported SAC Perform these actions 74 A Get command is issued but the sensor indicated that the location is empty. Look into the location (cell or drive) and see if a cartridge is present. If no cartridge is found in this location: • Perform the actions in Service Action Code 02. If a cartridge is found: • Contact ATAC. 78 A Put command is issued but the sensor indicated that the cartridge is not present in the Gripper Assembly.
Table 7-1 SAC Reporting Reported SAC E0 Perform these actions A cartridge is stuck in the Gripper Assembly Perform the following: • Turn off the main power Switch located on the lower panel of the rear door of all modules. • Open the front door of the module where the Accessor is located. • Lift the Accessor to the top of the Y axis rail and insert an allen wrench to stop the Accessor from sliding down. • Pull open the Insert/Eject Station.
Table 7-1 SAC Reporting Reported SAC E5 Perform these actions The SCSI bus connected to Library Port 0 is not properly terminated. Check the SCSI bus for termination. If the problem is not corrected or persists, contact ATAC. E6 The SCSI bus connected to Library Port 1 is not properly terminated. Check the SCSI bus for termination. If the problem is not corrected or persists, contact ATAC. E7 The Gripper Assembly detected that a cartridge is not present in the Grip fingers.
Operator Intervention Messages Table 7-2 describes the operator intervention messages. Each intervention message is composed of four lines. The third and/or fourth lines may contain variable information specific to the message. For SAC indications, refer to Service Action Codes on page 7-3. For Code error modifiers, refer to the Scalar 1000 Maintenance Guide. Table 7-2 Operator Intervention Messages Message Lines Alert not found. Press Enter. None WARNING! The accessor could not be parked.
Table 7-2 Operator Intervention Messages Message Lines 01 Aug 2000 Variables ->SYSTEM REBOOTEDTOO MANY ERRORS!
Table 7-2 Operator Intervention Messages Message Lines Variables INTERVENTION NEEDED! Storage may be full, or IE magazine types may be incorrect. None SERVICE MODE Cannot go online None Replace Terminator 7-12 Error Messages BAD MEDIA Cannot get type. Please label media. Cell %X (%Y %Z %A %B) %X %Y %Z %A %B = = = = = index number frame number section letter column number row number INTERVENTION NEEDED! Remove cartridge from cell.
Table 7-2 Operator Intervention Messages Message Lines 01 Aug 2000 Variables DRIVE COMM FAILED Please verify drive communication path. Drive %X (%Y %Z %A %B) %X %Y %Z %A %B = = = = = index number frame number section letter column number row number DRIVE SCSI ID FAILED Please verify drive SCSI id setting for drive %X (%Y %Z %A %B) %X %Y %Z %A %B = = = = = drive number frame number section letter column number row number BUTTON PUSH FAILED Please correct drive push-button state.
Table 7-2 Operator Intervention Messages Message Lines 7-14 Error Messages Variables POWER SUPPLY ERROR Check AC connection and 24V converter number %X %X = 24 Volt Connector Number POWER SUPPLY ERROR Check AC connection and 48V converter number %X %X = 48 Volt Connector Number POWER SUPPLY ERROR Check AC box and AC connection to primary/second DC supplies None INVALID LABEL Please check cart.
Error Log Codes In the event of an error, record the error information from the Error Log screen for any future reference. Refer to Error Log Dialog on page 5-35.
7-16 Error Messages 600561-A
Index -AAbout Dialog . . . . . . . . . . . . . . . . . . . . . . . . . 5-177 Accessor . . . . . . . . . . . . . . . . . . 2-3, 2-6, 2-12, 4-39 Accessor Submenu . . . . . . . . . . . . . . . . . . . . . 5-159 Advanced Dialog . . . . . . . . . . . . 5-71, 5-96, 5-175 Applying the Barcode Label . . . . . . . . . . . . . . 6-11 Area of Application . . . . . . . . . . . . . . . . . . . . . . 3-5 Assistance with problems . . . . . . . . . . . . . . . . . 1-5 Associated Documents . . . . . . . . . . . . . . . . .
Expansion Module . . . . . . . . . . . . . 2-3, 2-13, 4-37 external label . . . . . . . . . . . . . . . . . . 2-13, 2-15, 6-6 seventh character . . . . . . . . . . . . . . . 2-15, 6-7 -FFibre Channel . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17 Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7 Frames . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3, 2-8 Functional Description . . . . . . . . . . . 2-5, 2-6, 4-8 Functional Unit . . . . . . . . . . . . . . .
Security Dialog . . . . . . . . . . . . . . 5-84 Trace Dialog . . . . . . . . . . . . . . . . 5-83 Status Menu Selection . . . . . . . . . . . . . . . 5-12 Drives Submenu . . . . . . . . . . . . . . . . 5-37 Drive State Dialog . . . . . . . . . . . 5-38 Media Info Dialog . . . . . . . . . . . 5-40 Library Submenu . . . . . . . . . . . . . . . . 5-13 Element Dialog . . . . . . . . . . . . . . 5-28 Logs Submenu . . . . . . . . . . . . . . 5-32 SCSI Submenu . . . . . . . . . . . . . . 5-14 System Dialog . . .
Teach New Dialog . . . . . . . . . . . . . . . . . . . . . 5-167 Teach Submenu . . . . . . . . . . . . . . . . . . . . . . . 5-167 Time Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . 5-115 Trace Dialog . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-83 -UUp Arrow push button . . . . . . . . . . . . . . . . . . . 4-5 Update Microcode Dialog . . . . . . . . . . . . . . . 5-118 Upload Drives Dialog . . . . . . . . . . . . . . . . . . 5-125 Using the Operator Panel Common Help Buttons . . .