Specifications

Service Menu Selection 5-139
01 Aug 2000
Wrap Dialog
When
Wrap
is selected, the Wrap Dialog is displayed. See
Figure 5-154. This option tests the ability of the system sensors
to report state changes.
Figure 5-154 Wrap Dialog
Step 1 Press the button to select the current parameter
value and move to the next parameter.
or
Press the button or the button to toggle
between the parameter values in the following list.
Parameters Values
Device Indicates the list of sensors to
test (All, Gripper_Tach,
IE_Lock, Y_Home, X_Home,
Touch_Tip, Gripper_Open,
Retract_Complete, IE_Open)
Accept Y to accept changes
N to reject changes
Press the button to confirm the changes.
Step 2 Repeat Step 1 for all of the parameters in the list.
If
Y
is selected for the Accept parameter value, the
Response Screen appears. Refer to Figure 5-155 on page
5-140. Otherwise, the changed parameters continue to
display but no action is taken.
Device:
ALL SENSORS <
Results :
Accept: N