Specifications

Scalar 1000 Startup 4-37
01 Aug 2000
Scalar 1000 Startup
Use the following procedure to start the Scalar 1000.
With the RMU installed, Power Up must be delayed by at
least 10 seconds after a power down condition to allow the
RMU to properly shutdown.
Step 1 Ensure that all doors are closed and mechanically
locked on the Control Module and any Expansion
Modules.
Step 2 Locate the Main Power Switch at the lower right
corner in the back panel door of the Control
Module.
Step 3 Push in the top of the rocker Main Power Switch on
the Control Module. Repeat the process for any
Expansion Modules.
Movements of mechanical components in the Scalar 1000
can cause serious injury. Before turning the Main Power
Switch
ON
, confirm that no danger exists to personnel or
property.
Power is applied to the Scalar 1000. The Control Module
embedded software starts the boot process. The
(Power On) indicator illuminates and the Initial Screen
displays status.
The firmware initiates a self test procedure and an
auto-configuration and inventory operation. After all
firmware procedures are completed, the LCD screen on
the Operator Panel displays the Initial Screen. The Initial
Screen is illustrated by Figure 4-15.
Figure 4-15 Initial Screen
Caution
Warning
MAIN SCREEN
MM/DD/YY
Idle
Online.........Ready
HH/MM/SS